Skip to main content
  • Original research
  • Open access
  • Published:

Optimal imaging time points considering accuracy and precision of Patlak linearization for 89Zr-immuno-PET: a simulation study

Abstract

Purpose

Zirconium-89-immuno-positron emission tomography (89Zr-immuno-PET) has enabled visualization of zirconium-89 labelled monoclonal antibody (89Zr-mAb) uptake in organs and tumors in vivo. Patlak linearization of 89Zr-immuno-PET quantification data allows for separation of reversible and irreversible uptake, by combining multiple blood samples and PET images at different days. As one can obtain only a limited number of blood samples and scans per patient, choosing the optimal time points is important. Tissue activity concentration curves were simulated to evaluate the effect of imaging time points on Patlak results, considering different time points, input functions, noise levels and levels of reversible and irreversible uptake.

Methods

Based on 89Zr-mAb input functions and reference values for reversible (VT) and irreversible (Ki) uptake from literature, multiple tissue activity curves were simulated. Three different 89Zr-mAb input functions, five time points between 24 and 192 h p.i., noise levels of 5, 10 and 15%, and three reference Ki and VT values were considered. Simulated Ki and VT were calculated (Patlak linearization) for a thousand repetitions. Accuracy and precision of Patlak linearization were evaluated by comparing simulated Ki and VT with reference values.

Results

Simulations showed that Ki is always underestimated. Inclusion of time point 24 h p.i. reduced bias and variability in VT, and slightly reduced bias and variability in Ki, as compared to combinations of three later time points. After inclusion of 24 h p.i., minimal differences were found in bias and variability between different combinations of later imaging time points, despite different input functions, noise levels and reference values.

Conclusion

Inclusion of a blood sample and PET scan at 24 h p.i. improves accuracy and precision of Patlak results for 89Zr-immuno-PET; the exact timing of the two later time points is not critical.

Introduction

Therapeutic monoclonal antibodies (mAbs) are used in cancer treatment both in targeted therapy and in immunotherapy [1]. mAbs directly elicit their effect on their target or indirectly through mediation by the immune system. The effectiveness of this therapy is, however, patient specific and the therapy can cause serious side effects. Gaining more insight into the mechanisms of mAbs by tracking them inside the body may improve cancer treatment with mAbs.

Zirconium-89-immuno-positron emission tomography (89Zr-immuno-PET) allows visualization and quantification of the uptake of zirconium-89 labelled mAbs (89Zr-mAbs) in tumors and organs in vivo. The relatively long half-life of 89Zr is sufficient for imaging mAbs during the time they need to reach tissues [2]. Quantification of 89Zr-mAb uptake is commonly done using the standardized uptake value (SUV). SUV is defined as the activity concentration in a volume of interest, divided by the injected activity per unit of body weight [3]. Since SUV is a single value obtained from a single PET scan, SUV is not able to distinguish between non-specific 89Zr-mAb uptake in the blood or interstitial space volume fraction of the tissue, and specific uptake due to target engagement, unless either specific or non-specific uptake can be assumed to be negligible. In general, both non-specific and specific uptake contribute to the total uptake signal. Additionally, SUV considers only the injected activity and not the 89Zr-mAb plasma clearance over time [4].

An approach that does consider plasma activity concentrations for analyzing PET images is the use of compartment models [5]. Using a two-tissue compartment model assuming irreversible uptake of tracer, Patlak linearization can be applied [6]. A two-tissue irreversible compartment model is applicable to 89Zr-mAb uptake, because 89Zr residualizes in the tissue after mAb catabolism or target engagement [2]. The uptake of 89Zr-mAbs in tissue is quantified relative to the concentration of 89Zr-mAbs in blood plasma over time and therefore requires multiple blood samples and PET images. Since 89Zr-mAbs circulate in the body for several days [7], capturing the pharmacokinetics of 89Zr-mAbs requires multiple sampling days. However, minimizing the number of scans and samples is important in terms of patient safety and comfort. Selecting the optimal time points for blood sampling and PET imaging of 89Zr-mAbs is therefore crucial.

Patlak linearization provides several advantages over SUV. From Patlak linearization, reversible and irreversible 89Zr-mAb uptake can be quantified per volume of interest. Additionally, Patlak can potentially also distinguish between non-specific and specific 89Zr-mAb uptake, by comparing Patlak results to baseline Patlak values for tissues without target expression [8]. Moreover, Patlak linearization uses the measured plasma kinetics and thus takes variations in plasma clearance between subjects or at various mass doses into account. Yet, like SUV, Patlak linearization assumes that receptor availability or occupancy remains constant during the course of the PET studies and does not consider redistribution of cells or targets, as will be discussed later.

Previous research has applied Patlak linearization for quantifying 89Zr-mAbs uptake in patients [8, 9]. In these studies, PET scans were obtained two to four times between 2 and 192 h p.i. Blood was sampled up to five times on the day of injection and with every PET scan [8, 9]. This resulted in a maximum of three time points for Patlak linearization. The unavoidable sparse data sampling introduces uncertainties in the data which may affect Patlak results. Evaluating the magnitude of the effects of sparse data sampling will provide more information on the accuracy and precision of Patlak results.

In this study, the effect of imaging time points on the accuracy and precision of Patlak results was evaluated by means of simulations, including the following variables: different input functions (IFs), different noise levels for tissue activity curves (TACs) and tissues with different levels of reversible and irreversible uptake.

Methods

To study the effects of different time points on Patlak results, TACs were simulated using Patlak linearization, three time points were included, noise was added and Patlak values were calculated. These steps were repeated as a function of different variables.

Patlak linearization

Patlak linearization can be used to estimate the irreversible and reversible uptake of 89Zr-mAb in tissue based on graphical analysis of multiple-time tissue uptake data [6]. The analysis is based on a compartment model consisting of a reversible and an irreversible tissue compartment. The reversible tissue compartment represents 89Zr-mAb in the plasma and interstitial space of the tissue or reversible target binding, and reaches an equilibrium state after some time. The irreversible tissue compartment represents irreversible binding of 89Zr-mAb (e.g., non-specific catabolism or irreversible target binding). After equilibrium is reached, the activity concentration in tissue (ACt) is the sum of both parts. The reversible part is then proportional to the activity concentration in plasma (ACp) and the irreversible part is proportional to the area under the curve (AUC) of the ACp (AUCp), which is the integral of ACp (Eq. 1). Dividing both sides of Eq. 1 by ACp results in a linear relation known as the Patlak equation (Eq. 2) [6, 9]. The slope of this equation is Ki, which represents the nett rate of irreversible uptake [h −1]. Ki is a measure for the catabolic rate of tissue without target expression and a measure for both catabolic rate and target engagement of tissue with target expression [8]. The offset is the VT, the ratio between tissue and plasma concentration at equilibrium, which is related to the reversible part. (Eq. 2).

$${\text{AC}}_{t} = K_{i} \cdot {\text{AUC}}p + V_{T} \cdot {\text{AC}}_{p}$$
(1)
$$\frac{{{\text{AC}}_{t} }}{{{\text{AC}}_{p} }} = K_{i} \cdot \frac{{\mathop \smallint \nolimits_{0}^{t} {\text{AC}}_{P} \left( x \right){\text{d}}x}}{{{\text{AC}}_{P} }} + V_{T}$$
(2)

Multiple population IFs were obtained from literature as input for the ACp. A literature search for papers containing plasma/serum sampling data of 89Zr-mAb concentration in humans resulted in five papers as listed in Table 1. From these papers, the concentration 89Zr labelled mAbs in plasma/serum over time was obtained using PlotDigitizer (version 2.6.8, http://plotdigitizer.sourceforge.net/). The purpose of using IFs from literature was to use IFs that could be obtained in practice. Therefore, instead of using the raw data points, a bi-exponent (Eq. 3) was fitted through the data, see Fig. 1. The concentrations of the three IFs 89Zr-trastuzumab, 89Zr-pertuzumab and 89Zr-huJ591 were chosen as input for ACp in the simulations, as they presented three different clearance rates.

Table 1 Five papers provided 89Zr-mAb plasma/serum activity concentration data
Fig.1
figure 1

Plasma or serum activity concentrations in percentage injected activity per liter as a function of time in hours post-injection for 89Zr-huJ591 (191.3 ± 9 MBq, 25 mg) [10], 89Zr-trastuzumab (185 MBq, 50 mg) [11], 89Zr-pertuzumab (74 MBq, 20 or 50 mg) [12], 89Zr-DFO-MSTP2109A (184 MBq, 10 mg) [13] and 89Zr-AlbudAb (14 MBq, 1 mg) [14]. The bold lines represent the input functions used for the simulations, and the dashed lines represent the input functions not included in the simulations. %IA/L = percentage injected activity per liter, h p.i. = hours post-injection

The Patlak equation is used to simulate ACt as function of ACp, Ki and VT, i.e., to generate TACs. The given Ki and VT for generating the TAC are called ‘reference Ki (rKi)’ and ‘reference VT (rVT)’. The mathematical derivation for the TAC is as follows. ACp is described by a bi-exponential function (Eq. 3). AUCp can be obtained by integration of Eq. 3 between moment of injection and moment of PET scan, resulting in Eq. 4. Substitution of Eqs. 3 and 4 into Eq. 1 gives the equation for the TAC (ACt) as a function of rKi, rVT and coefficients of the bi-exponential equation of the IF (Eq. 5):

$${\text{AC}}_{p} = A \cdot {\text{e}}^{ax} + B \cdot {\text{e}}^{bx}$$
(3)
$${\text{AUC}}p = \mathop \smallint \limits_{0}^{t} \left( {A \cdot {\text{e}}^{ax} + B \cdot {\text{e}}^{bx} } \right){\text{d}}x = \frac{{A \cdot \left( {{\text{e}}^{ax} - 1} \right)}}{a} + \frac{{B \cdot \left( {{\text{e}}^{bx} - 1} \right)}}{b}$$
(4)
$${\text{AC}}_{t} = rK_{i} \cdot \left( {\frac{{A \cdot \left( {{\text{e}}^{ax} - 1} \right)}}{a} + \frac{{B \cdot \left( {{\text{e}}^{bx} - 1} \right)}}{b}} \right) + rV_{t} \cdot \left( {A\cdot{\text{e}}^{ax} + B \cdot {\text{e}}^{bx} } \right)$$
(5)

Sparse sampling and noise

For a given IF, rKi and rVT, values for ACp and ACt were determined with the equations above on three given time points, mimicking the sparse sampling in practice. AUCp was determined, but now by numerical integration of the IF, considering only four time points of ACp (see red line first panel Fig. 2). Additionally, noise was added to values for ACt at the given three time points. Standard deviations (SDs) of ACt were approximated based on counting statistics, which behaves as a Poisson distribution with SD≈√N and N is number of counts [15]. The SD at any given time point was approximated with Eq. 6, where the SD at t = 0 is predefined. Assuming equal scanning durations within a study, the ratio N(0):N(t) is assumed to be equal to the ratio between non-decay corrected activity concentrations ncACt(0):ncACt(t) (Eq. 7). To incorporate variability in the standard deviation, noise was added using the MATLAB function randn [16]. Subsequently, the percentage SD was calculated and applied on the decay corrected ACt for adding noise to ACt (Eq. 8).

$${\text{SD}}\left( t \right) = \frac{{{\text{SD}}\left( 0 \right)}}{{\sqrt {N\left( 0 \right)/N\left( t \right)} }}$$
(6)
$${\text{SD}}\left( t \right) = \frac{{{\text{SD}}\left( 0 \right)}}{{\sqrt {{\text{ncAC}}_{t} \left( 0 \right)/{\text{ncAC}}_{t} \left( t \right)} }}$$
(7)
$$AC_{t,noise} \left( t \right) = AC_{t} \left( t \right) + AC_{t} \left( t \right)*\% SD\left( t \right)*randn$$
(8)
Fig. 2
figure 2

Patlak linearization for 89Zr-pertuzumab input function and time activity curves with 5% noise, rKi = 1·10−3 [h−1], rVT = 0.2 and time points 24, 48 and 96 h post-injection. A: Activity concentrations in plasma (red) and tissue (blue), full curve based on reference values (rACp and rACt) and calculated based on simulations (sACp and sACt) in percentage injected activity per liter as a function of time p.i. B: Patlak plot; activity concentration in tissue (ACt) divided by activity concentration in plasma (ACp) as a function of area under the plasma curve (AUCp) divided by activity concentration in plasma (ACp). Based on reference values (black), simulated values (green) and linear regression of the simulated values (pink). %IA/L = percentage injected activity per liter

Variability in ACp as a result of counting statistics ranged from SD = 0.2–0.4%, based on previously in house counted blood samples. The noise in ACp was assumed to be negligible compared to the noise in the TAC and was not included in the simulations.

Patlak analysis of simulated TACs

Subsequently, Patlak linearization (Eq. 2) was applied on the generated ACp, AUCp and ACt with noise on the given time points, from which the slope (Ki) and offset (VT) could be determined, see Fig. 2. Simulations were repeated 1000 times to incorporate the effect of noise. The mean and standard error (SE) of the simulated Ki and VT were obtained to compare with rKi and rVT for evaluating bias and variability.

Performance of Patlak analysis

Accuracy and precision of Patlak results were evaluated as a function of the following variables: time points of evaluation, rKi and rVT, and noise level of ACt. Each simulation included a time point at 0 h p.i. for ACp. Additionally, three of the following time points in hours post-injection were considered: 24, 48, 96, 144 and 192, which resulted in 10 time point combinations. The chosen values for rKi were 1, 5 and 20 ∙10–3 h−1, representing real values of Ki for tissue without target expression [8], and two levels of target expression, respectively. The chosen rVT were 0.1, 0.2 and 0.5. These values were comparable to baseline values for VT as found by Jauw et al. [8], which agreed with predicted values for VT as sum of antibody biodistribution coefficient [17] and the plasma volume fraction. The noise levels of the TAC at time 0 were varied from 5%, 10% to 15%, equal to noise levels for the TAC previously used in a Patlak simulation study [18]. Simulations were performed in MATLAB (v9.3.0.713579) [16] using in-house written code (see Additional file 1).

Results

Simulations showed that bias in Ki was negative in all situations, see Figs. 3, 4 and 5 and Table 2. Inclusion of a time point at 24 h p.i. improved accuracy and precision of Patlak results in almost all simulations. Simulations with 89Zr-huJ591, 89Zr-trastuzumab and 89Zr-pertuzumab IF, noise level of 5%, rKi of 5·10–3 h−1 and rVT of 0.2 are shown in Fig. 3, and results are listed in Table 2. Including a time point at 24 h p.i. reduced bias and variability in VT for all three IF. Bias in Ki was reduced for 89Zr-huJ591 and remained similar for 89Zr-trastuzumab and 89Zr-pertuzumab. Variability in Ki remained similar for 89Zr-huJ591 and reduced slightly for 89Zr-trastuzumab and 89Zr-pertuzumab. Therefore, time point 24 h p.i. was included in all subsequent simulations.

Fig. 3
figure 3

Percentage bias and variability of Ki (A) and VT (B) per time point combination, for 89Zr-huJ591, 89Zr-trastuzumab and 89Zr-pertuzumab input functions and time activity curves with 5% noise, rKi = 5·10−3 [h−1] and rVT = 0.2. Combinations including 24 h post-injection showed smaller bias and variability than combinations without 24 h p.i. rKi = reference Ki, rVT = reference VT

Fig. 4
figure 4

Absolute Ki (A and B) and VT (C and D) values per time point combination for rKi = 1 and 20·10–3 [h−1], and for rVT = 0.1 and 0.5, for 89Zr-pertuzumab input function and time activity curves with 5% noise. All time point combinations on the x-axis also included 24 h post-injection. rKi = reference Ki, rVT = reference VT

Fig. 5
figure 5

Absolute Ki (A and B) and VT (C and D) values per time point combination for 89Zr-huJ591, 89Zr-trastuzumab and 89Zr-pertuzumab input functions, with rKi = 1 and 20·10−3 [h−1], rVT = 0.1 and time activity curves with 5% noise. All time point combinations on the x-axis also included 24 h post-injection. rKi = reference Ki, rVT = reference VT

Table 2 Smallest and largest percentage bias and variability of simulations with and without time point 24 h p.i. for both Ki and VT

Simulations with 89Zr-pertuzumab as IF and 5% noise level showed that bias in Ki ranged from − 0.5% (absolute bias of − 5·10–6 for Ki = 1·10−3 and Vt = 0.1) to − 6% (absolute bias of − 1.1·10−3 for Ki = 20·10−3 and Vt = 0.5) and bias in VT ranged from 2% (absolute bias of 0.01 for Vt = 0.5 and Ki = 1·10−3) to − 16% (absolute bias of − 0.016 for Vt = 0.1 and Ki = 1·10−3). Increasing the values for rKi and rVT resulted in increased variability in Ki and VT. Higher values for rKi also increased bias in Ki. However, bias in Ki resulting from increased rVT and bias in VT resulting from increased rKi and rVT remained similar, see Fig. 4.

Simulations with 89Zr-huJ591, 89Zr-trastuzumab and 89Zr-pertuzumab IF, rKi of 1·10−3 h−1 and rVT of 0.2 showed a threefold increase in variability in Ki and VT with higher noise levels, bias remained similar. For 89Zr-huJ591, increasing the noise level from 5 to 15% increased variability in Ki (SE from 23.0 to 68.0% and from 30.0 to 90.6%, respectively) and variability in VT (SE from 10.1 to 29.6% and 29.2 to 86.1%, respectively), while biases remained similar for Ki (from − 4.9 to − 5.1 and − 16 to − 16%, respectively) and VT (from − 1.6 to − 2.3% and 2.3 to 1.8%, respectively). Results of the other two IFs showed the same pattern. The noise level dependency was similar for higher rKi and rVT, however with higher bias and variability because of increased rKi and rVT.

A decrease in AUCp of the IF (in the order 89Zr-pertuzumab, 89Zr-trastuzumab, 89Zr-huJ591) resulted in increased bias in Ki and increased variability in VT with increased rKi, see Fig. 5. For rKi values of 20·10−3 h−1, bias in Ki also depended on the included time points, where the combinations 24, 48 and 192 h p.i. and 24, 144 and 192 h p.i. showed a larger underestimation of Ki of − 16% (absolute bias of − 3.2·10−3 for Ki = 20·10−3 and Vt = 0.1) for 89Zr-huJ591 IF as compared to − 10% for 89Zr-trastuzumab (absolute bias of − 2.0·10−3 for Ki = 20·10−3 and Vt = 0.1) and − 5.4% for 89Zr-pertuzumab IF (absolute bias of − 1.1·10−3 for Ki = 20·10−3 and Vt  = 0.1). Decreased AUCp of the IF also showed increased variability in Ki and VT for increased rVT; however, bias remained similar.

Overall, when including time point 24 h p.i., there were only small differences found in bias and variability between different time point combinations. Only for high Ki values and the 89Zr-huJ591 IF (with faster clearance of the 89Zr-mAb from blood), bias in Ki and VT showed a larger dependence on included time points, see Fig. 5. For all IFs, rKi, rVT, and time point combinations with noise level of 5%, percentage bias in Ki ranged from − 0.5 to − 16%.

Discussion

This study evaluated the effect of the choice of imaging time points on the accuracy and precision of Patlak linearization for 89Zr-immuno-PET, considering different conditions. Simulations showed that inclusion of a PET scan and blood sample at 24 h p.i. improves accuracy and precision of Patlak results. Different combinations of later time points did not change the accuracy and precision in most cases. Moreover, increase in rKi, rVT and noise level decreased accuracy and precision of Patlak results. Additionally, IFs with smaller AUCp showed decreased accuracy and precision of Patlak results as compared to IFs with larger AUCp.

Underestimation of K i

Bias in Ki was negative in all simulations. This can be explained by the shape of the IF in combination with the calculation of AUCp in the Patlak equation [6]. In case the IF is fully described, for instance with a bi-exponential equation, determining the AUCp by integration will result in the true value for AUCp. However, when only a finite set of points is known from the IF, determining the AUCp will be based on trapezoidal numerical integration. For the simulations in this study, the latter applies, because data sampling is always finite. Since the activity concentration in plasma decreases over time in an exponential manner, the shape of the IF is curved downwards, leading to an overestimation of the AUCp with trapezoidal numerical integration. The overestimated AUCp increases the x-coordinates of the Patlak plot, which is AUCp/ACp, while the y-coordinates remain the same, because the ratio ACt/ACp does not change. This results in a decreased positive slope of the Patlak plot, e.g., negative bias of Ki.

24 h time point

Inclusion of time point 24 h p.i. showed to improve accuracy and precision of Patlak linearization. This is also due to the better assessment of the shape of the IF and the calculation of AUCp as detailed before. The better the curve of the IF is described, by adding a time point in the most curved part of the IF, the more accurate the determination of AUCp and Patlak parameters. One assumption for Patlak linearization is that equilibrium is reached between the 89Zr-mAb concentration in plasma and in the reversible tissue compartment, meaning that all fluxes are constant with respect to time [6]. In this study, activity concentrations in tissue were simulated by means of Patlak linearization and therefore were directly in equilibrium with activity concentrations in plasma. However, mAbs are relatively large proteins, therefore distribution inside the body takes relatively long, so tissue is not in rapid equilibrium with plasma [7]. Therapeutic antibodies cetuximab and trastuzumab showed approximately homogeneous distributions after 24 h p.i. in tumor-bearing mice [19]. For this reason, a period of 24 h was estimated to reach equilibrium between tissue and plasma. Additionally, from a practical point of view, it would not be possible to include time points after approximately 12 h, because PET scans should then be obtained outside working hours. Hence, time points before 24 h p.i. were not included in the simulations. This moment of equilibrium may differ between 89Zr-mAbs, and inclusion of a slightly earlier or later time point may be better depending on the mAb pharmacokinetics.

Time point combinations

After inclusion of the 24 h p.i. time point, different time point combinations barely influenced Patlak results, which is advantageous from a practical perspective. Postponing a late imaging time point to a different day would not influence Patlak results. This is in contrast with obtaining the SUV, for which differences in the uptake time between injection and PET scan does influence the result, because SUV changes as a function of time [20]. In case the assumption of equal clearance between patients is true, comparisons of SUVs between patients would only be possible for PET scans that are obtained at the same uptake time post-injection [4]. Therefore, postponing a PET scan, resulting in different scan days for patients accompanied by different plasma activity concentrations, will influence SUV results. Apart from the ability to distinguish between reversible and irreversible, and potentially between non-specific and specific uptake of 89Zr-mAbs [8], the option to postpone a PET scan is another advantage of using Patlak linearization over using SUV in the quantification of 89Zr-immuno-PET.

Reference K i and V T

Simulations showed that increasing rKi and rVT resulted in similar or increased bias and variability in both Ki and VT. As Patlak linearization is only applied when the assumption of irreversible uptake is met, Ki is never zero. Additionally, Jauw et al. [8] showed that organs without target expression have Ki values higher than zero, representing the catabolic rate of 89Zr-mAbs in healthy tissue. Values for Ki in this study are therefore all above zero.

Noise levels

In this study, noise was approximated based on counting statistics, which resulted in noise increasing over time. This was similar to results from a study about noise-induced variability in PET imaging for 89Zr-immuno-PET, where recovery coefficients (RC) also increased over time from day 0 to day 6 [21]. RC was defined as 1.96*SD(%). RCs found for Kidney, lung, spleen and liver combined ranged from 2 to 11 [21], resulting in a maximum SD of approximately 5%. Similarly, SD derived from the RCs of tumor SUVpeak results in 15%. Simulations including TACs with a 5% noise level may therefore represent biodistribution and TACs with a 15% noise level may represent tumor uptake. Increasing the noise level from 5 to 15% only increased the variability, biases remained the same. Additionally, results of simulations with a noise level of 15% showed the same pattern as simulations with a 5% noise level and were chosen not to be presented.

Input functions

The literature search provided five different 89Zr-mAb plasma IFs in patients, of which three were used for the simulations, while there are currently 119 therapeutic antibodies approved by the FDA [22]. However, these three 89Zr-mAb plasma IFs used in this study provide a wide range of clearances, covering substantial variability in IFs.

Simulations showed a dependency of Patlak results on the IF. For high rKi, accuracy and precision in Patlak results decreased with AUC of the IF (i.e., faster clearance), in the following order: 89Zr-pertuzumab, 89Zr-trastuzumab and 89Zr-huJ591. A decrease in AUCp will result in lower x-coordinates of the Patlak plot, thereby bringing the datapoints closer together resulting in higher contribution of noise. The AUCp is the integral of the activity concentration in plasma, which is the total 89Zr-mAbs present in the plasma cumulated over time from injection to moment of PET scan. For the simulations, the IF and rKi were regarded as two independent variables; however, they are physiologically related. For IFs with lower AUCp, so faster clearance, higher irreversible uptake in tissue (rKi) is expected. However, simulations showed that a higher rKi for the 89Zr-huJ591 IF resulted in decreased accuracy of Ki (− 16%) and precision of VT as compared to the other IFs. This indicates that accuracy and precision of Patlak results are worse for 89Zr-mAbs with faster clearance combined with higher irreversible uptake. However, for volumes of interest showing high irreversible uptake, a bias in Ki of − 16% would not change the (clinical) decision-making based on the data, because the observed irreversible uptake would still be high.

This study considers input functions with binding of targets on cells that do not redistribute during the course of the PET studies (HER2 for trastuzumab and pertuzumab, and PSMA for huJ591). However, the usefulness of Patlak linearization may be limited in case of 89Zr-mAbs that bind to mobile immune cells, such as the PD-1 receptors on T-cells. In order to apply Patlak linearization, an equilibrium between reversible processes is assumed as well as a constant density of specific targets or receptors. Changes in receptor availability during the course of the study may introduce inaccuracies in Patlak linearization. Yet, Patlak analysis also has several advantages over SUV. Patlak linearization can also be applied with higher mass dose. However, there are two phenomena that need to be considered. First of all, higher mass doses will result in slower plasma clearance. Patlak linearization takes into account the mAb concentration in plasma (or input function) and no assumptions are required with regard to (changes in) plasma clearance as the measured plasma kinetics are used. Secondly, a higher administered mass dose will result in lower uptake in tissue of interest. Patlak linearization is still valid with higher mass doses; however, lower Ki values are expected because of the reduced receptor availability/higher receptor or target occupancy. Also, Menke-van der Houven van Oordt et al. [9] showed in their study that Patlak linearization applied to PET imaging data with different administered mass doses allows evaluation of the optimal therapeutic dose. By plotting the Patlak Ki values against increasing mass doses a S-curve can be obtained. Ki values decrease because of target binding competition between labeled and unlabeled mAbs. This curve allows evaluation of the 50% inhibitory mass dose (ID50). The ID50, the dose at which 50% of the targets are occupied, can be used in establishing the optimal therapeutic dose [9].

Conclusion

This study evaluated the effect of imaging time points on the accuracy and precision of Patlak results, for different IFs, imaging time points, noise levels, and tissues with different levels of reversible and irreversible uptake. Quantification of 89Zr-immuno-PET using Patlak linearization can generate accurate results within − 0.5% and − 16% bias for Ki (at a 5% noise level), provided that a 24 h p.i. time point and two later time points are included. The exact timing of the two other scans and samples is, however, not critical as opposed to SUV-based quantification.

Availability of data and materials

All data and scripts generated during the current study are available from the corresponding author on reasonable request.

References

  1. Kimiz-Gebologlu I, Gulce-Iz S, Biray-Avci C. Monoclonal antibodies in cancer immunotherapy. Mol Biol Rep. 2018;45:2935–40.

    Article  CAS  Google Scholar 

  2. van Dongen G, Beaino W, Windhorst AD, et al. The role of (89)Zr-immuno-PET in navigating and derisking the development of biopharmaceuticals. J Nucl Med. 2021;62:438–45.

    Article  Google Scholar 

  3. Boellaard R. Standards for PET image acquisition and quantitative data analysis. J Nucl Med. 2009;50(Suppl 1):11S-20S.

    Article  CAS  Google Scholar 

  4. Lammertsma AA, Hoekstra CJ, Giaccone G, Hoekstra OS. How should we analyse FDG PET studies for monitoring tumour response? Eur J Nucl Med Mol Imaging. 2006;33(Suppl 1):16–21.

    Article  Google Scholar 

  5. Vriens D, Visser EP, de Geus-Oei LF, Oyen WJ. Methodological considerations in quantification of oncological FDG PET studies. Eur J Nucl Med Mol Imaging. 2010;37:1408–25.

    Article  Google Scholar 

  6. Patlak CS, Blasberg RG, Fenstermacher JD. Graphical evaluation of blood-to-brain transfer constants from multiple-time uptake data. J Cereb Blood Flow Metab. 1983;3:1–7.

    Article  CAS  Google Scholar 

  7. Lobo ED, Hansen RJ, Balthasar JP. Antibody pharmacokinetics and pharmacodynamics. J Pharm Sci. 2004;93:2645–68.

    Article  CAS  Google Scholar 

  8. Jauw YWS, O’Donoghue JA, Zijlstra JM, et al. (89)Zr-Immuno-PET: toward a noninvasive clinical tool to measure target engagement of therapeutic antibodies in vivo. J Nucl Med. 2019;60:1825–32.

    Article  CAS  Google Scholar 

  9. der Houven M, van Oordt CW, McGeoch A, Bergstrom M, et al. Immuno-PET imaging to assess target engagement: experience from (89)Zr-Anti-HER3 mAb (GSK2849330) in patients with solid tumors. J Nucl Med. 2019;60:902–9.

    Article  Google Scholar 

  10. Pandit-Taskar N, O’Donoghue JA, Beylergil V, et al. (8)(9)Zr-huJ591 immuno-PET imaging in patients with advanced metastatic prostate cancer. Eur J Nucl Med Mol Imaging. 2014;41:2093–105.

    Article  CAS  Google Scholar 

  11. O’Donoghue JA, Lewis JS, Pandit-Taskar N, et al. Pharmacokinetics, biodistribution, and radiation dosimetry for (89)Zr-trastuzumab in patients with esophagogastric cancer. J Nucl Med. 2018;59:161–6.

    Article  CAS  Google Scholar 

  12. Ulaner GA, Lyashchenko SK, Riedl C, et al. First-in-human human epidermal growth factor receptor 2-targeted imaging using (89)Zr-Pertuzumab PET/CT: dosimetry and clinical application in patients with breast cancer. J Nucl Med. 2018;59:900–6.

    Article  CAS  Google Scholar 

  13. O’Donoghue JA, Danila DC, Pandit-Taskar N, et al. Pharmacokinetics and biodistribution of a [(89)Zr]Zr-DFO-MSTP2109A anti-STEAP1 antibody in metastatic castration-resistant prostate cancer patients. Mol Pharm. 2019;16:3083–90.

    Article  CAS  Google Scholar 

  14. Thorneloe KS, Sepp A, Zhang S, et al. The biodistribution and clearance of AlbudAb, a novel biopharmaceutical medicine platform, assessed via PET imaging in humans. EJNMMI Res. 2019;9:45.

    Article  Google Scholar 

  15. Cherry SR, Sorenson J, Phelps ME, Methé BM. Physics in nuclear medicine. Med Phys. 2004;31:2370–1.

    Article  Google Scholar 

  16. MATLAB. version 9.3.0.713579 (R2017b). Natick, Massachusetts: The MathWorks Inc.; 2017.

  17. Shah DK, Betts AM. Antibody biodistribution coefficients: inferring tissue concentrations of monoclonal antibodies based on the plasma concentrations in several preclinical species and human. mAbs. 2013;5:297–305.

    Article  Google Scholar 

  18. van Sluis J, Yaqub M, Brouwers AH, Dierckx R, Noordzij W, Boellaard R. Use of population input functions for reduced scan duration whole-body Patlak (18)F-FDG PET imaging. EJNMMI Phys. 2021;8:11.

    Article  Google Scholar 

  19. Lee CM, Tannock IF. The distribution of the therapeutic monoclonal antibodies cetuximab and trastuzumab within solid tumors. BMC Cancer. 2010;10:255.

    Article  Google Scholar 

  20. Keyes JW Jr. SUV: standard uptake or silly useless value? J Nucl Med. 1995;36:1836–9.

    PubMed  Google Scholar 

  21. Jauw YWS, Heijtel DF, Zijlstra JM, et al. Noise-induced variability of immuno-PET with zirconium-89-labeled antibodies: an analysis based on count-reduced clinical images. Mol Imaging Biol. 2018;20:1025–34.

    Article  CAS  Google Scholar 

  22. Cai HH. Therapeutic Monoclonal Antibodies Approved by FDA in 2020. Clin Res Immunol. 2021;4(1):1–2.

Download references

Acknowledgements

Not applicable

Funding

This work has received funding from the Innovative Medicines Initiative 2 Joint Undertaking (JU) under grant agreement No. 831514 (Immune-Image). The JU receives support from the European Union’s Horizon 2020 research and innovation programme and EFPIA.

Author information

Authors and Affiliations

Authors

Contributions

All authors contributed to the study conception and design. Data collection and analysis were performed by JW, MH and RB. The first draft of the manuscript was written by JW and all authors (MH, JP, WM, YJ and RB) commented on previous versions of the manuscript. All authors (JW, MH, JP, WM, YJ and RB) read and approved the final manuscript.

Corresponding author

Correspondence to Jessica E. Wijngaarden.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1.

In-house written MATLAB code for Patlak linearization. The in-house written MATLAB function provided in Supplemental 1 was used for Patlak linearization calculations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Wijngaarden, J.E., Huisman, M.C., Pouw, J.E.E. et al. Optimal imaging time points considering accuracy and precision of Patlak linearization for 89Zr-immuno-PET: a simulation study. EJNMMI Res 12, 54 (2022). https://doi.org/10.1186/s13550-022-00927-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13550-022-00927-6

Keywords