- Original research
- Open access
- Published:
Region- and voxel-based quantification in human brain of [18F]LSN3316612, a radioligand for O-GlcNAcase
EJNMMI Research volume 11, Article number: 35 (2021)
Abstract
Background
Previous studies found that the positron emission tomography (PET) radioligand [18F]LSN3316612 accurately quantified O-GlcNAcase in human brain using a two-tissue compartment model (2TCM). This study sought to assess kinetic model(s) as an alternative to 2TCM for quantifying [18F]LSN3316612 binding, particularly in order to generate good-quality parametric images.
Methods
The current study reanalyzed data from a previous study of 10 healthy volunteers who underwent both test and retest PET scans with [18F]LSN3316612. Kinetic analysis was performed at the region level with 2TCM using 120-min PET data and arterial input function, which was considered as the gold standard. Quantification was then obtained at both the region and voxel levels using Logan plot, Ichise's multilinear analysis-1 (MA1), standard spectral analysis (SA), and impulse response function at 120Â min (IRF120). To avoid arterial sampling, a noninvasive relative quantification (standardized uptake value ratio (SUVR)) was also tested using the corpus callosum as a pseudo-reference region. Venous samples were also assessed to see whether they could substitute for arterial ones.
Results
Logan and MA1 generated parametric images of good visual quality and their total distribution volume (VT) values at both the region and voxel levels were strongly correlated with 2TCM-derived VT (r = 0.96–0.99) and showed little bias (up to − 8%). SA was more weakly correlated to 2TCM-derived VT (r = 0.93–0.98) and was more biased (~ 16%). IRF120 showed a strong correlation with 2TCM-derived VT (r = 0.96) but generated noisier parametric images. All techniques were comparable to 2TCM in terms of test–retest variability and reliability except IRF120, which gave significantly worse results. Noninvasive SUVR values were not correlated with 2TCM-derived VT, and arteriovenous equilibrium was never reached.
Conclusions
Compared to SA and IRF, Logan and MA1 are more suitable alternatives to 2TCM for quantifying [18F]LSN3316612 and generating good-quality parametric images.
Introduction
Neurofibrillary tangles—aggregates of hyperphosphorylated tau protein—are a typical feature of Alzheimer's disease (AD) and other tauopathies [1]. The process of O-GlcNAcylation [2] involves attaching O-linked β-N-acetylglucosamine (O-GlcNAc) at the serine and threonine residue of tau protein [3]. Two enzymes modulate this process: O-GlcNAc transferase and O-GlcNAc hydrolase (O-GlcNAcase, OGA), which catalyze the attachment and detachment of O-GlcNAc, respectively. Because this process modifies the key protein involved in tauopathies, and because OGA is located on chromosome 10q24.1 (a locus associated with late-onset AD [4, 5]), interest has grown in O-GlcNAcylation as a therapeutic target. Notably, O-GlcNAcylation hinders abnormal phosphorylation and aggregation of tau, thus stabilizing the microtubule-associated tau protein [6, 7]. Indeed, tau-specific and overall O-GlcNAc are decreased in the brains of individuals with AD [8, 9]. Preclinical studies have also demonstrated that upregulation of O-GlcNAcylation by OGA inhibitors reduces pathologic tau phosphorylation and aggregation and prevents neurodegeneration [10,11,12]. These findings suggest that OGA inhibition may be a potential strategy for treating tauopathies [13, 14].
Our laboratory recently characterized [18F]LSN3316612 (fluorine-18-N-(5-(((2S,4S)-2-methyl-4-(6-fluoropyridin-2-yloxy) piperidin-1-yl)methyl)thiazol-2-yl)acetamide), a positron emission tomography (PET) radioligand capable of measuring OGA in brain. An initial in vitro study demonstrated that [18F]LSN3316612 was a highly specific and potent OGA ligand [15] with good imaging properties in nonhuman primates [16]. Human studies subsequently confirmed that [18F]LSN3316612 was an excellent PET radioligand for imaging OGA in vivo. Specifically, brain uptake was well quantified as total distribution volume (VT) using the two-tissue compartment model (2TCM); VT measures showed low variability and moderate reliability under test and retest conditions, and VT was stable about two hours after injection, suggesting no significant radiometabolite accumulation in brain [17]. However, only VT derived from compartment modeling was examined at the region level, and 2TCM is not suitable for voxel-wise analysis due to its high computational demand [18]. Thus, the feasibility of other kinetic models remains unknown, particularly with regard to generating high-quality parametric images of [18F]LSN3316612 binding. The present study used the previously obtained test–retest dataset [17] to identify which kinetic model(s) yield binding parameters that are close to 2TCM-derived VT, which is considered the gold standard for quantifying [18F]LSN3316612 binding.
Materials and methods
Participants
The current study reanalyzed data from 10 healthy volunteers (five males, five females; 43 ± 11 years) who had enrolled in a previous protocol (NCT03632226) [17] and underwent both test and retest PET scans with [18F]LSN3316612. The National Institutes of Health (NIH) Combined Neurosciences Institutional Review Board approved the original study (Protocol 16-M-0105), and informed consent was obtained from all participants.
Imaging studies
Details regarding the participants, PET study protocol and processing, and measurement of [18F]LSN3316612 in plasma are described in detail in the previously published paper [17]. Briefly, all participants underwent dynamic PET scans using an mCT scanner (Siemens Medical Solutions, Cary, NC, USA) after intravenous administration of [18F]LSN3316612. Data were reconstructed into 45 frames (6 × 0.5 min, 3 × 1 min, 2 × 2 min, 34 × 5 min) using a three-dimensional ordered subset expectation maximization algorithm. For the retest study, participants were scanned on a separate day (median interval = 49 ± 48 days (range 8–150)).
For structural magnetic resonance (MR) imaging, sagittal T1-weighted brain MR images were obtained in all participants using a 3 T Philips Achieva scanner (Philips Healthcare; Andover, MA) with turbo field echo sequence (repetition time = 8.1 ms, echo time = 3.7 ms, flip angle = 8, matrix = 181 × 256 × 256, voxel size = 1.000 × 0.983 × 0.983 mm).
Image preprocessing—such as motion correction, MR segmentation, coregistration between PET and MR, and atlas normalization—was performed using the PNEURO pipeline implemented in PMOD 3.903 (PMOD, Zurich, Switzerland). A total of 83 predefined regions-of-interest (ROIs) from the Hammers N30R83 maximum probability atlas [19] were adjusted to the MRIs of individual participants and subsequently combined into an individual template comprising 16 consolidated regions. For region-level analyses, brain time-activity curves were obtained by applying the individual ROI template on the dynamic PET images transformed into MR space. For voxel-level analyses, parametric images were obtained with the PXMOD tool, and the average value over each brain region was calculated by overlaying the same Hammers atlas used for the region-level analyses onto the parametric images.
Arterial blood was concomitantly withdrawn during the brain PET scan. Whole and plasma concentrations of [18F]LSN3316612 were measured using an automatic gamma counter and corrected after radiometabolites were separated using high-performance liquid chromatography (HPLC). Both whole-blood and total plasma activity curves were then fitted to a tri-exponential function, and the plasma parent fraction was fitted using a Hill function [20]. Parent radioactivity in plasma, which was implicitly generated by the product of the total plasma activity curve and parent fraction in PMOD, was used as the input function.
Kinetic models
The acquisition lasted for 180Â min; however, only the first 120Â min of PET and blood data were used for kinetic evaluation because our previous study had demonstrated that 120Â min of brain and plasma data were sufficient to measure the VT of [11C]LSN3316612 with small bias and good identifiability [17].
Requiring blood input function
Because the 2TCM is superior to the one-tissue compartment model for quantifying [18F]LSN3316612 binding [17], the gold standard for the present study was total VT calculated with 2TCM. Brain activity was corrected for its vascular components, assuming that blood volume was 5% of total brain volume [21]. The delay in radioligand arrival between the radial artery and the brain was corrected by fitting the whole-brain time-activity curve.
Regional VT values were also obtained with the Logan plot (Loganvoi), Ichise's multilinear analysis-1 (MA1voi), and spectral analysis (SAvoi). In addition to VT, SA was used to calculate the impulse response function at 120Â min (IRF120) (see Appendix).
Similar to region-level analyses, voxel-wise parametric images were created using the following four methods: Logan plot (Loganvoxel), MA1 (MA1voxel), SA (SAvoxel), and IRF (IRF120voxel). For Loganvoxel and MA1voxel, starting PET frames used for regression were selected based on t* of the whole brain. Regional binding parameters—VT and IRF120—were then obtained by measuring the average value of the voxels in each region.
No requirement for input function
For relative measurement of [18F]LSN3316612 binding, time-averaged PET images at 20-min intervals were created starting from 60 min (60–80, 80–100, and 100–120 min). Standardized uptake value (SUV) was measured for each ROI, and regional SUV ratio (SUVR) was then calculated by dividing the SUV of each region by that of the corpus callosum. Because OGA is known to be ubiquitously expressed throughout the human brain, a suitable reference region devoid of specific binding does not exist. Therefore, the corpus callosum, which has the lowest VT, was designated as a normative region for SUV. Among the three time intervals (60–80, 80–100, and 100–120 min), the 100–120-min data showed the best correlation between SUVR and 2TCM-derived VT values.
Assessing the arteriovenous equilibrium of [18F]LSN3316612
Venous blood samples were simultaneously obtained in four participants, with arterial sampling at 30, 60, 120, and 180Â min. Plasma parent concentration of [18F]LSN3316612 was obtained for each of the four timepoints and compared with those obtained from the arterial blood samples. Whether arterial blood samples could be replaced by venous blood samples was explored based on the agreement of the plasma parent concentrations at those four timepoints.
Statistical analysis
For all participants, the VT value obtained with 2TCM at the region level was considered as the gold standard [17]. Pearson’s coefficient (r) and percentage bias (calculated as the mean relative difference from 2TCM estimates) were used to assess the correlation and difference between each binding parameter and VT calculated from 2TCM, respectively. For the voxel-level analysis, the mean value of each brain region obtained from the parametric images was compared to the corresponding 2TCM-derived VT using the same brain atlas.
VT comparison among different kinetic models at both region and voxel levels was performed by repeated measures analysis of variance (ANOVA) followed by a Fisher’s least significant difference (LSD) test.
Test–retest variability and absolute test–retest variability were calculated as follows:
To assess test–retest reliability, the intra-class correlation coefficient (ICC) of each region was calculated as follows [22]:
where BSMSS and WSMSS are the mean sums of squares between subjects and within subjects, respectively. ICC values between 0.50 and 0.75 were defined as moderate, between 0.75 and 0.90 as good, and above 0.9 as excellent [23].
Statistical significance was set at P < 0.05, and all statistical analyses were conducted using GraphPad Prism 5 (GraphPad Software, La Jolla, CA, USA) and SPSS 25 (IBM Corp., New York, USA).
Results
Kinetic model comparison
At the region level, the whole-brain VT obtained with Loganvoi, MA1voi, and IRF120voi were strongly correlated with VT obtained with 2TCM (r > 0.95, P < 0.001 for all) (Table 1). SAvoi showed a significant but slightly weaker correlation (r = 0.90, P < 0.001).
In contrast, SUVR calculated using a pseudo-reference region did not correlate with 2TCM-derived VT (r = − 0.24, P = 0.497) and was thus excluded from further analyses. A pseudo-reference region is a region that has specific binding but is relatively unaffected by the disease [24, 25]. Loganvoi and MA1voi slightly underestimated VT values; the bias relative to VT obtained with 2TCM was − 4.8% (95% confidence interval (CI), − 5.5 to − 4.1%; range − 3.0 to − 8.3%) for Loganvoi and − 8.3% (95% CI − 9.3 to − 7.4%; range − 6.2 to − 13.5%) for MA1voi (Table 2). In contrast, SAvoi overestimated VT with a bias of 11.6% (95% CI, 10.8 to 12.5%; range 7.7 to 13.9%). However, repeated measures ANOVA determined that the differences in VT were not statistically significant among 2TCM, Loganvoi, MA1voi, and SAvoi (F(3, 36) = 2.340, P = 0.090).
At the voxel level, Logan, MA1, and SA generated parametric images of good quality that represented OGA density in the brain (Fig. 1). The parametric images generated by IRF120voxel were noisier, although the overall image quality was acceptable for quantification. Whole-brain VT values obtained with Loganvoxel, MA1voxel, SAvoxel, and IRF120 were strongly correlated with VT obtained with 2TCM (r > 0.95, P < 0.001 for all; Table 1). Loganvoxel and MA1voxel showed a small bias in VT with 2TCM: − 7.3% (95% CI − 9.2 to − 5.5%; range − 5.6 to − 20.4%) for Loganvoxel and − 2.4% (95% CI − 3.6 to − 1.2%; range − 7.3% to 1.6%) for MA1voxel (Table 2). SAvoxel overestimated VT with a bias of 15.5% (95% CI, 14.6 to 16.4%; range 9.3 to 17.3%). The repeated measures ANOVA was significant (F(3, 36) = 3.206, P = 0.034) and Fisher’s LSD test showed that SA differed from both Logan and MA1. Indeed, SA tended to overestimate VT, whereas graphical analyses underestimate VT compared to compartmental modeling.
Although both test and retest datasets were analyzed separately in the same way, only the test data are reported here because the retest results were very similar (data not shown).
Test–retest variability and reliability
VT obtained with Logan, MA1, and SA showed low test–retest variability and moderate test–retest reliability, similar to those obtained with 2TCM-derived VT. At the region level, the absolute test–retest variability ranged from 10.7 to 11.5%, and ICC ranged from 0.69 to 0.73, compared to 11.7% and 0.70 for 2TCM, respectively (Table 3). At the voxel level, the absolute test–retest variability ranged from 10.4% to 12.0%, and ICC ranged from 0.58 to 0.80; these values demonstrate moderate to good reliability. In contrast, IRF120 showed significantly worse test–retest variability (19.7% to 20.8%) than the other techniques at both the region and voxel levels, though it had comparable ICC (0.71).
Use of venous input function
Equilibrium was not reached between arterial and venous blood data measured at four timepoints. At 30 min, arterial blood samples showed higher plasma concentrations than venous blood samples, with an average arterial-venous difference of 14.3 ± 12.2% (range 7.3 to 32.5%). Plasma concentrations were almost equal at 60 min, with an average difference of − 1.5 ± 5.9% (range − 8.0 to 0.0%) between arterial and venous blood data; however, at later timepoints, the average difference became greater (− 5.5 ± 14.0% (range − 13.5 to 15.4%) at 120 min and − 25.0 ± 21.0% (range − 54.7 to − 6.7%) at 180 min (Fig. 2).
Discussion
The present study—which used previous data to explore different kinetic model(s) for quantifying [18F]LSN3316612 binding—found that Logan and MA1 were the models whose VT estimations were the closest to those obtained using the gold standard of 2TCM, particularly with regard to generating good-quality parametric images. Compared to SA, Logan and MA1 showed the smallest bias of VT compared to 2TCM. Although VT values from all techniques were strongly correlated with 2TCM-derived VT at both the region and voxel levels, the biases for Logan and MA1 were generally within − 8% of the 2TCM-derived VT values, while SA showed a larger bias (~ 16%). In contrast, while IRF120 was strongly correlated with 2TCM-derived VT in both region- and voxel-wise analyses, its test–retest variability was significantly higher than for the other techniques. Therefore, IRF120 should not be considered a viable alternative for [18F]LSN3316612 quantification. Finally, the lack of correlation between SUVR and 2TCM-derived VT values and the absence of arteriovenous equilibrium 180 min post-injection suggest that arterial sampling is needed to quantify [18F]LSN3316612 scans.
Previous studies from our laboratory showed that [18F]LSN3316612 is suitable for clinical research [17]. Specifically, brain uptake was high and could be well-quantified as VT with 2TCM. Furthermore, VT was stable over time and showed low variability and fair reliability under test–retest conditions. However, quantifying increased OGA density during disease development and progression requires not only favorable kinetic properties but also validation of an appropriate analytic methodology. In this context, voxel-based analyses may enable early detection compared to region-based analyses. Theoretically, region-based analyses may be appropriate if a study has an a priori hypothesis about a specific brain region. However, voxel-based analyses may identify changes in smaller and isolated areas that take place during early disease stages. Indeed, a change in a subset of voxels may be smoothed out when all voxel values are averaged together to a large ROI. Moreover, voxel-based analyses examine the whole brain without a priori spatial assumption and are more appropriate for explorative studies of radioligands that bind across all brain regions, such as [18F]LSN3316612. Voxel-based analyses are more affected by noise, and adjacent voxels are not statistically independent observations because they are spatially correlated due to the finite resolution of the scanner. In our study, however, the impact of noise at the voxel level was minimized by merging the individual voxels of the parametric images into relatively noiseless large regions.
Although 2TCM is considered to be the gold standard for quantifying [18F]LSN3316612 in the brain, it is unsuitable for voxel-level analyses. The conventional nonlinear least-squares fitting is susceptible to the high noise encountered at the voxel level [26]. As an example, Rizzo and colleagues [27] found that generating parametric images of [11C](R)-rolipram with 2TCM took about 20Â h of computing time per participant and resulted in a map where more than half of voxels had to be discarded because of outlying values. Parametric images may also be generated using the basis function derived from 2TCM [28], but the results may be inaccurate and are still computationally challenging [18]. Alternative methods were recently proposed for computing all rate constants of 2TCM within a reasonable time and improving robustness to measurement noise, but they require validation in other reversible radioligands [18, 29].
In this context, the present study found that Logan and MA1 were valid alternatives to 2TCM at the region level and suitable for generating parametric images. Logan and MA1 use a linear estimator, thus enabling the rapid estimation of macro-parameters, such as VT [30, 31]. Moreover, these graphical approaches do not require a priori knowledge of compartmental configurations. One drawback of the graphical analyses is that they underestimate VT due to noise in the PET data, especially for Logan [32]. In our study, however, the bias was both small (up to about 8%) and similar between region-based and voxel-based analyses.
Interestingly, SA similarly uses a linear estimator and does not depend on specific compartment model configuration, thus making it both computationally rapid and widely applicable to PET radioligand research. However, in the present study, SA overestimated VT by about 16%. SA is well known to be sensitive to noise in the data [33], with a tendency to overestimate macro-parameters like VT, especially for shorter acquisition times.
Reliable quantification of [18F]LSN3316612 requires arterial blood sampling. To eliminate the need for this invasive and burdensome procedure, the present study examined use of a pseudo-reference region and also measured simultaneous arterial and venous concentrations in four participants. Because pseudo-reference regions are relatively unaffected by the disease [32, 33], this approach can only be properly validated using groups of both patients and healthy controls [34]. However, given the very poor correlation with VT already present in healthy participants, a pseudo-reference region may not be a viable option for [18F]LSN3316612. Not only was a suitable arteriovenous equilibrium never reached over 180Â min, but considerable inter-subject variability was also observed. However, this lack of suitable equilibrium between the artery and vein is common for radioligands [35]. Even if an average equilibrium is obtained at a late timepoint, considerable inter-subject variability will result in VT estimation errors in one or more participants. Given the often limited number of participants in PET studies, the presence of these outliers could easily compromise results.
Finally, and as previously reported [17], VT was generally higher in the second scan compared to the first, regardless of the method used. If this were purely due to chance, the probability of obtaining nine higher retest scans from 10 subjects would be smaller than 1%. Curiously, Coughlin and colleagues similarly found a systematic higher VT value in retest scans compared to test scans when scanning six subjects with the TSPO radioligand [11C]DPA-713 [36]. In their study—and in contrast to ours—all retest scans were acquired on the same day as the test scan. The authors speculated that these findings might be due to factors such as hormone-mediated phasic changes in TSPO expression, tonic changes due to stress/anxiety related to the procedure, or alterations in blood cholesterol levels due to food intake between the scans. As yet unidentified physiological variables may also explain our results.
Conclusions
Logan plot and MA1 showed the smallest VT bias when compared with 2TCM at the region and voxel levels and generated good-quality parametric images. Therefore, both models were better than SA or IRF for quantifying [18F]LSN3316612 uptake in human brain.
Availability of data and materials
The data that support the findings of this study will be made available, without restriction, by request to Dr. Robert Innis (innisr@mail.nih.gov). The data are not publicly available due to privacy and ethical restrictions.
Abbreviations
- 1TCM:
-
One-tissue compartment model
- 2TCM:
-
Two-tissue compartment model
- ANOVA:
-
Analysis of variance
- ICC:
-
Intra-class correlation coefficient
- IRF120 :
-
Impulse response function calculated at 120Â min
- MA1:
-
Ichise's multilinear analysis-1
- MR:
-
Magnetic resonance
- O-GlcNAc:
-
O-Linked β-N-acetylglucosamine
- OGA:
-
O-GlcNAc hydrolase (O-GlcNAcase)
- PET:
-
Positron emission tomography
- ROI:
-
Region of interest
- SA:
-
Standard spectral analysis
- SUV:
-
Standardized uptake value
- SUVR:
-
Standardized uptake ratio between the region/voxel and pseudo-reference region
- TRV:
-
Test–retest variability
- V T :
-
Total distribution volume
References
Lee VM, Goedert M, Trojanowski JQ. Neurodegenerative tauopathies. Annu Rev Neurosci. 2001;24:1121–59.
Hart GW, Kreppel LK, Comer FI, et al. O-GlcNAcylation of key nuclear and cytoskeletal proteins: reciprocity with O-phosphorylation and putative roles in protein multimerization. Glycobiology. 1996;6:711–6.
Hart GW, Housley MP, Slawson C. Cycling of O-linked β-N-acetylglucosamine on nucleocytoplasmic proteins. Nature. 2007;446:1017–22.
Gong CX, Liu F, Iqbal K. O-GlcNAcylation: A regulator of tau pathology and neurodegeneration. Alzheimers Dement. 2016;12:1078–89.
Grupe A, Li Y, Rowland C, et al. A scan of chromosome 10 identifies a novel locus showing strong association with late-onset Alzheimer disease. Am J Hum Genet. 2006;78:78–88.
Yuzwa SA, Macauley MS, Heinonen JE, et al. A potent mechanism-inspired O-GlcNAcase inhibitor that blocks phosphorylation of tau in vivo. Nat Chem Biol. 2008;4:483–90.
Yu Y, Zhang L, Li X, et al. Differential effects of an O-GlcNAcase inhibitor on tau phosphorylation. PLoS ONE. 2012;7:e35277.
Forster S, Welleford AS, Triplett JC, Sultana R, Schmitz B, Butterfield DA. Increased O-GlcNAc levels correlate with decreased O-GlcNAcase levels in Alzheimer disease brain. Biochim Biophys Acta. 2014;1842:1333–9.
Liu F, Iqbal K, Grundke-Iqbal I, Hart GW, Gong CX. O-GlcNAcylation regulates phosphorylation of tau: a mechanism involved in Alzheimer’s disease. Proc Natl Acad Sci USA. 2004;101:10804–9.
Yuzwa SA, Shan X, Jones BA, et al. Pharmacological inhibition of O-GlcNAcase (OGA) prevents cognitive decline and amyloid plaque formation in bigenic tau/APP mutant mice. Mol Neurodegener. 2014;9:42.
Graham DL, Gray AJ, Joyce JA, et al. Increased O-GlcNAcylation reduces pathological tau without affecting its normal phosphorylation in a mouse model of tauopathy. Neuropharmacology. 2014;79:307–13.
Yuzwa SA, Shan X, Macauley MS, et al. Increasing O-GlcNAc slows neurodegeneration and stabilizes tau against aggregation. Nat Chem Biol. 2012;8:393–9.
Dong L, Shen S, Chen W, et al. Discovery of novel inhibitors targeting human O-GlcNAcase: docking-based virtual screening, biological evaluation, structural modification, and molecular dynamics simulation. J Chem Inf Model. 2019;59:4374–82.
Macauley MS, Vocadlo DJ. Increasing O-GlcNAc levels: An overview of small-molecule inhibitors of O-GlcNAcase. Biochimica et Biophysica Acta (BBA) General Subjects. 2010;1800:107–21.
Lu S, Haskali MB, Ruley KM, et al. PET ligands [18F]LSN3316612 and [11C]LSN3316612 quantify O-linked-beta-N-acetyl-glucosamine hydrolase in the brain. Sci Transl Med. 2020;12:543.
Paul S, Haskali MB, Liow JS, et al. Evaluation of a PET radioligand to image O-GlcNAcase in brain and periphery of rhesus monkey and knock-out mouse. J Nucl Med. 2019;60:129–34.
Lee JH, Liow JS, Paul S, et al. PET quantification of brain O-GlcNAcase with [18F]LSN3316612 in healthy human volunteers. EJNMMI Res. 2020;10:20.
Kudomi N, Maeda Y, Hatakeyama T, Yamamoto Y, Nishiyama Y. Fully parametric imaging with reversible tracer 18F-FLT within a reasonable time. Radiol Phys Technol. 2017;10:41–8.
Hammers A, Allom R, Koepp MJ, et al. Three-dimensional maximum probability atlas of the human brain, with particular reference to the temporal lobe. Hum Brain Mapp. 2003;19:224–7.
Gunn RN, Sargent PA, Bench CJ, et al. Tracer kinetic modeling of the 5-HT1A receptor ligand [carbonyl-11C]WAY-100635 for PET. Neuroimage. 1998;8:426–40.
Huisman MC, van Golen LW, Hoetjes NJ, et al. Cerebral blood flow and glucose metabolism in healthy volunteers measured using a high-resolution PET scanner. EJNMMI Res. 2012;2:63.
Shrout PE, Fleiss JL. Intraclass correlations: uses in assessing rater reliability. Psychol Bull. 1979;86:420–8.
Koo TK, Li MY. A Guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med. 2016;15:155–63.
Turkheimer FE, Selvaraj S, Hinz R, et al. Quantification of ligand PET studies using a reference region with a displaceable fraction: application to occupancy studies with [11C]-DASB as an example. J Cereb Blood Flow Metab. 2012;32:70–80.
Salinas CA, Searle GE, Gunn RN. The simplified reference tissue model: model assumption violations and their impact on binding potential. J Cereb Blood Flow Metab. 2015;35:304–11.
Maltz JS. Parsimonious basis selection in exponential spectral analysis. Phys Med Biol. 2002;47:2341–65.
Rizzo G, Veronese M, Zanotti-Fregonara P, Bertoldo A. Voxelwise quantification of [11C](R)-rolipram PET data: a comparison between model-based and data-driven methods. J Cereb Blood Flow Metab. 2013;33:1032–40.
Hong YT, Fryer TD. Kinetic modelling using basis functions derived from two-tissue compartmental models with a plasma input function: general principle and application to [18F]fluorodeoxyglucose positron emission tomography. Neuroimage. 2010;51:164–72.
Castellaro M, Rizzo G, Tonietto M, et al. A Variational Bayesian inference method for parametric imaging of PET data. Neuroimage. 2017;150:136–49.
Logan J, Fowler JS, Volkow ND, et al. Graphical analysis of reversible radioligand binding from time-activity measurements applied to [N-11C-methyl]-(-)-cocaine PET studies in human subjects. J Cereb Blood Flow Metab. 1990;10:740–7.
Ichise M, Toyama H, Innis RB, Carson RE. Strategies to improve neuroreceptor parameter estimation by linear regression analysis. J Cereb Blood Flow Metab. 2002;22:1271–81.
Slifstein M, Laruelle M. Effects of statistical noise on graphic analysis of PET neuroreceptor studies. J Nucl Med. 2000;41:2083–8.
Cunningham VJ, Gunn RN, Byrne H, Matthews JC. Suppression of noise artifacts in spectral analysis of dynamic PET data. In: Carson RE, Daube-Witherspoon ME, Herscovitch P, editors. Quantitative functional brain imaging with positron emission tomography. San Diego: Academic Press; 1998. p. 329–34.
Lyoo CH, Ikawa M, Liow JS, et al. Cerebellum can serve as a pseudo-reference region in Alzheimer disease to detect neuroinflammation measured with PET radioligand binding to translocator protein. J Nucl Med. 2015;56:701–6.
Zanotti-Fregonara P, Chen K, Liow J-S, Fujita M, Innis RB. Image-derived input function for brain PET studies: many challenges and few opportunities. J Cereb Blood Flow Metab. 2011;31:1986–98.
Coughlin JM, Wang Y, Ma S, Yue C, Kim PK, Adams AV, et al. Regional brain distribution of translocator protein using [11C]DPA-713 PET in individuals infected with HIV. J Neurovirol. 2014;20(3):219–32.
Veronese M, Rizzo G, Turkheimer FE, Bertoldo A. SAKE: a new quantification tool for positron emission tomography studies. Comput Methods Programs Biomed. 2013;111:199–213.
Weeks RA, Cunningham VJ, Piccini P, Waters S, Harding AE, Brooks DJ. 11C-diprenorphine binding in Huntington’s disease: a comparison of region of interest analysis with statistical parametric mapping. J Cereb Blood Flow Metab. 1997;17:943–9.
Tadokoro M, Jones A, Cunningham V, et al. Parametric images of 11C diprenorphine binding using spectral analysis of dynamic PET images acquired in 3D. In: Uemura K, Jones T, Lassen NA, Kanno I, editors. Quantification of Brain Function: Tracer Kinetics and Image Analysis in Brain PET. New York: Excerpta Medica; 1993. p. 289–94.
Acknowledgements
The authors thank the NIH Clinical Center PET Department (Chief: P. Herscovitch) for supplying the fluorine-18; Eli Lilly & Co. for providing precursors and support throughout the human study; PMOD Technologies for providing the image analysis software and support; and Ioline D. Henter (NIMH) for invaluable editorial assistance.
Funding
This work was supported jointly by the Intramural Research Program of the National Institute of Mental Health, National Institutes of Health (ZIAMH002793 and ZIAMH002852) and a Cooperative Research and Development Agreement with Eli Lilly & Co.
Author information
Authors and Affiliations
Contributions
JHL performed the data analysis and drafted the manuscript. MV provided the MATLAB codes for standard spectral analysis and impulse response function calculation. JSL coordinated the study and carried out data acquisition and analysis. CLM, MBH, and VWP contributed to radiosynthesis and quality control of [18F]LSN3316612. JAMS and SSZ analyzed blood samples and interpreted the data. RBI and PZF contributed to the study design and data interpretation. All authors read, revised, and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
The National Institutes of Health Combined Neurosciences Institutional Review Board approved this study (Protocol 16-M-0105), and informed consent was obtained from all participants.
Consent for publication
Not applicable.
Competing interests
All authors have no conflicts of interest to disclose, financial or otherwise.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Appendix
Appendix
Logan plot. The Logan plot is a model-independent graphical method for reversible tracers [30]. It performs a linearization of the data so that, after a certain time (t > t*), the slope can be related to the VT, according to the following equation:
where CT(t) is the tissue time-activity curve, CP(t) is the metabolite-corrected arterial input function, the slope (VT) is the total distribution volume, and b is the intercept, which becomes constant for t > t*.
Ichise's multilinear analysis. The multilinear analysis (MA1) is a modification of the Logan plot aimed at minimizing the bias induced by noise in the measurements [31]:
where CT(t) represents the tissue time-activity curve, CP(t) the metabolite-corrected arterial input function, VT the total distribution volume, and b the intercept of the Logan plot that becomes constant after an equilibration time (t > t*).
Spectral analysis. This technique is based on a single time-invariant input/single output model used to identify tissue kinetic components [37]. SA does not require prior knowledge of the number of compartments in the system. In SA, IRF represents the hypothetical concentration of the radioligand that would be measured after ideal instantaneous bolus injection. The tissue concentration, CT(t), is modeled as a convolution of the radiometabolite-corrected arterial input function, CP(t), with the sum of M + 1 distinct decreasing exponential function [37]:
where αj and βj are functions of the transfer rate constants assumed to be real-valued and nonnegative.
VT was calculated from the estimated spectrum as follows:
In the present study, the βj grid was defined with a maximum value of 250 and a logarithmic distribution of βj, j = 1, 2… M, using a spectral range from 0.005 to 5 min−1.
In addition, impulse response function at 120Â min (IRF120) was used as a proxy of the binding of [18F]LSN3316612 [38, 39]. SA was performed by using the source codes of SAKE on MATLAB R2016b (MathWorks, Natick, MA, USA) [37].
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Lee, JH., Veronese, M., Liow, JS. et al. Region- and voxel-based quantification in human brain of [18F]LSN3316612, a radioligand for O-GlcNAcase. EJNMMI Res 11, 35 (2021). https://doi.org/10.1186/s13550-021-00780-z
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s13550-021-00780-z