Skip to main content
  • Original research
  • Open access
  • Published:

The use of a proposed updated EARL harmonization of 18F-FDG PET-CT in patients with lymphoma yields significant differences in Deauville score compared with current EARL recommendations

Abstract

Background

The Deauville score (DS) is a clinical tool, based on the comparison between lesion and reference organ uptake of 18F-fluorodeoxyglucose (FDG), used to stratify patients with lymphoma into categories reflecting their disease status. With a plethora of positron emission tomography with computed tomography (PET-CT) hard- and software algorithms, standard uptake value (SUV) in lesions and reference organs may differ which affects DS classification and therefore medical treatment. The EANM Research Ltd. (EARL) harmonization program from the European Association of Nuclear Medicine (EANM) partly mitigates this issue, but local preferences are common in clinical practice. We have investigated the discordance in DS calculated from patients with lymphoma referred for 18F-FDG PET-CT reconstructed with three different algorithms: the newly introduced block-sequential regularization expectation-maximization algorithm commercially sold as Q. Clear (QC, GE Healthcare, Milwaukee, WI, USA), compliant with the newly proposed updated EARL recommendations, and two settings compliant with the current EARL recommendations (EARLlower and EARLupper, representing the lower and upper limit of the EARL recommendations).

Methods

Fifty-two patients with non-Hodgkin and Hodgkin lymphoma were included (18 females and 34 males). Segmentation of mediastinal blood pool and liver were semi-automatically performed, whereas segmentation of lesions was done manually. From these segmentations, SUVmax and SUVpeak were obtained and DS calculated.

Results

There was a significant difference in DS between the QC algorithm and EARLlower/EARLupper (p < 0.0001 for both) but not between EARLlower and EARLupper (p = 0.102) when SUVmax was used. For SUVpeak, there was a significant difference between QC and EARLlower (p = 0.001), but not for QC vs EARLupper (p = 0.071) or EARLlower vs EARLupper (p = 0.102). Five non-responders (DS 4–5) for QC were classified as responders (DS 1–3) when EARLlower/EARLupper was used, both when SUVmax and SUVpeak were investigated.

Conclusion

Using the proposed updated EARL recommendations compared with the current recommendations will significantly change DS classification. In select cases, the discordance would affect the choice of medical treatment. Specifically, the current EARL recommendations were more often prone to classify patients as responders.

Background

Over the years, there have been multiple advances in positron emission tomography with computed tomography (PET-CT) regarding both hard- and software. The new developments, such as introduction of time-of-flight, point-spread-function, smaller voxels, respiratory gating, silicon (Si) photomultiplier (PM) detectors and block-sequential regularized expectation maximization (BSREM) reconstruction algorithms (commercially sold as Q. Clear (QC), GE Healthcare, Milwaukee, WI, USA), have all contributed to a better image quality, improved small lesion detectability and more accurate quantification of radiopharmaceutical uptake [1].

In patients with Hodgkin and non-Hodgkin lymphoma, 18F-fluorodeoxyglucose (FDG) PET-CT has become the standard procedure in the staging, monitoring and restaging of disease. During therapy assessment at mid-treatment and after completion of chemotherapy, the Deauville score (DS) is recommended to discriminate between responders and non-responders [2, 3]. DS is a 5-point scale where the lesion with the most intense uptake is compared to the physiological uptake in the mediastinal blood pool and the liver. Responders are usually defined as DS 1–3 and non-responders as DS 4–5 [4,5,6].

The new developments in PET have been shown to affect the maximum standardized uptake value (SUV) in lesions [7]. It can therefore be suspected that a patient being examined on different PET-CT scanners with different hard- and software as well as with different acquisition parameters might receive different DS. To overcome this issue, the EANM Research Ltd. (EARL) harmonization programme from the European Association of Nuclear Medicine (EANM) has set up recommendations on how to perform PET imaging for oncologic purpose, including harmonization of the patient preparation, scan acquisition, image processing and interpretation of images, in order to be able to compare results from different PET-CT scanners [8]. This harmonization, however, does not take the most modern applications of PET hard- and software in consideration and might underestimate DS in small lesions. Recently, there has been a proposal for updating the EANM/EARL recommendations to include modern PET-CT equipment [9].

The aim of this study was to investigate whether using a novel state-of-the-art SiPM-based PET-CT with QC reconstruction (that complies with the newly proposed updating of the EANM/EARL recommendation) may affect DS compared with reconstructions meeting the current EANM/EARL harmonizing standard in patients with lymphoma, regarding both SUVmax and SUVpeak.

Methods

Patients

In this retrospective study, 57 patients who underwent clinical 18F-FDG PET-CT between November 2017 and March 2018 or August 2018 to October 2018 at Skåne University Hospital in Lund or Malmö, Sweden, were included initially. Patients admitted for baseline PET, mid-treatment (interim) PET (i-PET), end-of-treatment PET (EoT-PET) and suspicion of recurrence were included. Four patients did not have any discernible lesion on CT and one patient had a history of lymphoma but not at the time of the examination, which was performed for other reasons. These five patients were excluded, leaving 52 patients in the study.

PET-CT acquisition and reconstruction parameters

Three Discovery MI (GE Healthcare, Milwaukee, WI, USA) PET-CT systems were used for image acquisition. The systems were configured with four rings of detector blocks with lutetium yttrium oxyorthosilicate crystals coupled to an array of SiPM. The PET-detector has a transaxial field of view of 70 cm, an axial field of view of 20 cm and an overlap of 24% between bed positions. The sensitivity, according to NEMA standards, was 13 cps/kBq. The PET system was combined with a 128 slice CT.

All patients received an intravenous injection of 4 MBq/kg body weight of 18F-FDG with an accumulation time of 60 min before imaging and after at least 4 h of fasting and a glucose level ≤ 10 mM. If no contraindications existed, the patients were administered with beta-blockers before the examination. Patients were scanned from the inguinal region to the base of the skull. Acquisition time was 1.5 min per bed position. CT images were acquired for attenuation correction and anatomic correlation of the PET images. A diagnostic CT with intravenous and oral contrast or a low-dose CT without contrast was performed. In our clinical routine, a low-dose is performed if a previous diagnostic CT has been performed within 4 weeks. For diagnostic CTs, tube current modulation was applied by adjusting the tube current for each individual with a noise index of 42.25 and a tube voltage of 100 kV. For low-dose CT, the tube voltage was 120 kV with a noise index of 45. If a diagnostic CT was performed, it was used for attenuation correction (delayed venous phase of intravenous contrast). The same CT was used for attenuation correction for all PET reconstructions. The adaptive statistical iterative reconstruction technique (ASiR-V) was applied for all CT reconstructions.

In order to compare the reconstruction algorithms, we reconstructed different data series, where the selected reconstruction parameters were based on phantom measurements in accordance with the EARL standard [10]. The EARL standard defines lower and upper limits for the resolution recovery coefficient (RRC) for different sized spheres in the NEMA-phantom and limits of the noise level. Two reconstructions were made corresponding to the lower (EARLlower) and upper level (EARLupper) of the RRCs. The ordered subset expectation maximization (OSEM) algorithm was used without resolution recovery or time of flight. For the upper level, the images were reconstructed with 4 iterations, 16 subsets and a Gaussian post filter with a FWHM of 5 mm. For the lower-level images, the reconstructions were performed with 3 iterations, 8 subsets and a post-filter with 7 mm FWHM. A new EARL standard has been proposed where the RRC limits have been substantially increased to accommodate modern systems [9]. One reconstruction was made which yields EARL results that fall near the upper level of the new EARL standard. The QC reconstruction algorithm was used, with a beta value of 500 [11]. The slice thickness for all three reconstructions was 2.79 mm, the matrix and pixel size were 192 × 192 and 3.64 mm for the EARL reconstructions and 256 × 256 and 2.73 mm for the QC reconstruction.

Image analysis

A machine learning method described previously [12] was used to segment the liver and the mediastinal blood pool (thoracic part of the aorta) in the CT images. One radiology resident and one specialist in radiology and nuclear medicine corrected the automated segmentations when needed. Focal lesions within the liver were not included in the segmentation. The segmentations were then eroded by 3 voxels in all directions in order to avoid the edges of the respective organs. The EARL reconstructions were regridded to the same pixel size as the QC reconstruction prior to the erosion operation, giving an erode kernel of 8.2 × 8.2 × 8.4 mm for all reconstructions.

Lymphoma lesions were manually segmented in the CT images by the two physicians described above. The PET image could be overlaid to help segmentation in case of registration mismatch between the CT and PET, wherein such cases segmentations were slightly outside the CT lesion in order to include the lesion SUVmax.

The SUVmax and SUVpeak of the liver, blood pool and lesions were calculated using the segmentations made in the CT image translated to the corresponding locations in the PET images. Lesion SUVmax and SUVpeak were then compared to SUVmax and SUVpeak in the liver and blood pool in order to assign a DS.

Statistical analysis

Continuous patient parameters are presented as mean ± standard deviation (SD) and range and categorical variables as a percentage (%). A Friedman test comparing the DS obtained from the three different reconstruction algorithms was performed for SUVmax and SUVpeak, respectively. Significant p value was set at p < 0.05. Post hoc analysis with Wilcoxon signed-rank test was conducted with a Bonferroni correction applied, resulting in a significance level set at p < 0.0167. All statistical tests were performed using IBM SPSS version 25 (IBM, Armonk, NY, USA).

Results

Patients

Fifty-two patients with lymphoma were enrolled (16 Hodgkin’s lymphoma, 1 Burkitt’s lymphoma, 2 B cell lymphoma (subtype unavailable), 23 diffuse large B cell lymphoma, 6 follicular lymphoma, 2 mantle cell lymphoma, 1 peripheral T cell lymphoma and 1 anaplastic large cell lymphoma). There were in total 10 baseline PET, 13 i-PET, 26 EoT-PET examinations and 5 suspicion of recurrence. Two patients underwent both i-PET and EoT-PET, resulting in n = 54 PET-CT examinations. Patients were aged between 17 and 83 years of which 35% were women. The mean (± SD) weight was 79 ± 16 kg (range 46–137 kg) and the mean BMI was 25.8 ± 4.6 (range 16.9–39.6). The mean administrated 18F-FDG was 4.0 ± 0.15 MBq/kg (range 3.0–4.3 MBq/kg) and the mean accumulation time was 62 ± 4 min (range 54–78 min).

Quantitative analysis

None of the patients were classified as DS 1. Table 1 shows the number of patients classified as DS 2–5 for the three different reconstruction methods using SUVmax and SUVpeak.

Table 1 Classifications of DS for the three different reconstruction methods using SUVmax (A–C) and SUVpeak (D–F)

For SUVmax calculations, the Friedman test resulted in p < 0.0001. Five (9.3%) QC non-responders (DS 4–5) became responders (DS 2–3) with EARLlower and/or EARLupper. When comparing EARLlower with EARLupper, one patient changed from DS 3 to DS 4 and one patient changed from DS 4 to DS 3. No other patient changed from non-responder to responder or vice versa. Discordance in DS occurred in 18 cases (33.3%) when comparing QC with EARLlower, in 17 cases (31.5%) when comparing QC with EARLupper, and in 6 cases (11.1%) when comparing EARLlower with EARLupper. Discordant lesions were consistently downscaled in DS with either EARLlower or EARLupper compared to QC, except for one patient that was upscaled from DS 2 (QC) to DS3 (EARLupper). Wilcoxon tests resulted in QC vs EARLlower p < 0.0001, QC vs EARLupper p < 0.0001 and EARLlower vs EARLupper p = 0.102 (not significant).

For SUVpeak calculations, the Friedman test resulted in p = 0.003. Five (9.3%) QC non-responders became responders with EARLlower and/or EARLupper. When comparing EARLlower with EARLupper, one patient changed from DS 3 to DS 4 and one patient changed from DS 4 to DS 3. No other patient changed from non-responder to responder or vice versa. Discordance in DS occurred in 11 cases (20.4%) when comparing QC with EARLlower, in 15 cases (27.8%) when comparing QC with EARLupper and in 6 cases (11.1%) when comparing EARLlower with EARLupper. Discordant lesions were consistently downscaled in DS with either EARLlower or EARLupper compared to QC, except for four patients. Two patients were upscaled from DS 2 (QC) to DS3 (EARLupper), one patient from DS 3 (QC) to DS 4 (EARLupper) and one patient from DS 4 (QC) to DS 5 (EARLupper). Wilcoxon tests resulted in QC vs EARLlower p = 0.001, QC vs EARLupper p = 0.071 (not significant) and EARLlower vs EARLupper p = 0.102 (not significant).

Figure 1 shows an example of a patient with major discordance between QC, EARLlower and EARLupper DS. Figure 2 shows an example of a patient with good concordance for DS between the different reconstruction settings. Figure 3 shows the ratios between concordance, discordance and major discordances when comparing the reconstruction algorithms pairwise.

Fig. 1
figure 1

Major discordance in DS. Representative example of patient who had major discordance between QC, EARLlower and EARLupper DSs. This patient was classified as DS 5 on QC images, DS 3 on EARLlower images and DS 4 on EARLupper images. Transversal images in the thorax (aorta and lesion (indicated with an arrow) with highest SUVmax)—upper row—and in the abdomen (liver)—lower row—for the three different reconstructions including the corresponding CT are shown

Fig. 2
figure 2

Good concordance in DS. Representative example of patient who had good concordance between QC, EARLlower and EARLupper DSs. This patient was classified as DS 3 on all reconstruction settings. Transversal images in the thorax (aorta and lesion (indicated with an arrow) with highest SUVmax)—upper row—and in the abdomen (liver)—lower row—for the three different reconstructions including the corresponding CT are shown

Fig. 3
figure 3

Diagrams of pairwise comparisons between reconstruction algorithms. Numbers in coloured semi-circles represent n cases that had concordance, discordance and major discordance respectively when comparing reconstruction algorithms pairwise. For SUVmax, post hoc analysis with Wilcoxon signed-rank test with Bonferroni correction, QC compared to Elower and Eupper respectively yielded significant p value but not for Elower compared to Eupper. For SUVpeak, only QC compared to Elower yielded significant p value

Discussion

The latest proposed update to the EARL recommendations accommodates modern PET technologies, in particular time-of-flight and point spread function (PSF) [6]. Our study compares datasets that are compatible with both current and the proposed update to the EARL recommendations. The introduction of time-of-flight and PSF have shown to have minimal effect on the liver and mediastinal uptake [10], but new hardware and reconstruction algorithms provide higher SUV in small lesions [4], and therefore affect DS classification. Our findings show that if the proposed update to the EARL recommendations is accepted, it will have an impact on DS and therapy response evaluation. The studies behind the recommendations on DS and treatment evaluation are performed on older generations of PET-CT scanners, and it is not known whether the DS obtained from new state-of-the-art PET-CT scanners will have an impact on patient outcome in large cohorts.

Investigation of whether the choice of reconstruction method affects DS has recently been performed by Enilorac et al. [13], comparing one dataset with unfiltered PSF (Siemens HD) and one where a 6-mm Gaussian filter was applied to PSF images to match the EARL requirements. The proportion of major discordances was comparable to our findings for SUVmax but our conclusions differ. In their study of 126 patients, no difference in progression-free survival and overall survival was seen depending on the reconstruction method, when patients were classified as responders or non-responders. However, they analysed i-PET and EoT-PET separately, yielding small groups with only a few patients classified differently depending on the reconstruction method.

There are different aspects on DS that should be considered, such as how SUV in reference organs are measured, the cut-off for DS 5 and how to handle patients with a higher SUV in the mediastinum compared with liver. In our study, we used automatic segmentation of liver and mediastinum [6]. The edges around the liver and aortic wall were automatically truncated to avoid uptake from adjacent structures and the vessel wall. Segmentations were manually corrected when needed. This method increases the likelihood of obtaining the true SUVmax. When manual ROIs are placed in reference organs, there is an apparent risk of missing the true SUVmax. In a couple of the patients, we found the mediastinal SUVmax to be higher than the liver SUVmax. This was confirmed with manual ROI measurements (data not shown). There is no support in the literature or guidelines on how these cases should be managed in terms of DS classification. However, in our study, no patient had lesion uptake that was between uptake in the mediastinum and in the liver. There is no consensus where the cut-off point should be for DS 5, and both a limit of two or three times the maximum uptake in the liver has been proposed [4]. We classified DS 5 as two times the maximum uptake in the liver. There were few major discordances in DS (i.e. when a non-responder (QC) is reclassified as responder (EARLlower/EARLupper)) between reconstruction methods, which has clinical significance in terms of treatment strategy. If a worst-case scenario is preferred, then using settings that adhere to the newly proposed EARL recommendation is more suitable.

We included baseline exams in order to increase the study population, although DS is normally not calculated in baseline examinations. In theory, a follow-up scan could look like a baseline scan. In the retrospective analysis, baseline exams did not show any major discordances: for SUVmax, there were 2 discordances, and for SUVpeak, there were also 2 discordances. If all baseline scans were removed from the study, the results would show an even higher percentage of discordances across all pairwise comparisons between reconstruction algorithms.

No solid recommendation of how to obtain DS exists, although SUVmax appears to be the most commonly used method. In this study, we investigated the use of both SUVmax and SUVpeak, as proposed both by Barrington et al. [4] and the newly proposed EARL recommendations [9]. SUVmax is more noise dependent [14]; therefore, SUVpeak is a more stable measure. This was also true in our study, where we did not find any significant differences in DS between QC and EARLupper. However, SUVpeak requires a lesion of more than 1 cm in order to be relevant. SUVmax, on the other hand, has been shown to be unreliable in sub-centimetre lesions when PSF is used [15]. There is no standard definition of SUVpeak calculation which may be seen in differing implementations of SUVpeak calculations in various software. A harmonization across vendors is desirable to further increase its reproducibility.

Limitations

In this study, we included all patients with lymphoma, regardless of the indication for the PET-CT examination. In clinical routine, DS is only used for therapy assessment and not for initial staging/baseline. However, in order to increase the number of patients and the range of included DS, also, patients referred for baseline PET-CT were included. Despite this, we recognize the limitation of the study due to its small sample size and its monocentric nature.

Although we have showed considerable differences in DS between the reconstruction algorithms, it remains to be proven which reconstruction algorithm has the most favourable outcome for the patients. The type of lymphoma and the intensity of stage-adapted chemotherapy adds further complexity to the outcome.

1It would be of interest to compare the upper and lower limits of the newly proposed EARL recommendations, but for our PET-CT system, longer acquisition times are necessary to reach the new upper limit, which was not feasible for the current study.

Conclusions

There is a significant difference in DS classification when comparing the proposed update to EARL recommendations and the current recommendations. In select cases, the discordance would affect the choice of medical treatment. Specifically, the current EARL recommendations were more often prone to classify patients as responders compared with the recently proposed EARL update. Further studies are needed in order to prove which reconstruction algorithm is suitable for assessing patient outcome.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

ASiR-V:

Adaptive statistical iterative reconstruction technique

BSREM:

Block-sequential regularized expectation maximization

DS:

Deauville score

EANM:

European Association of Nuclear Medicine

EARL:

EANM Research Ltd.

EoT-PET:

End-of-treatment PET

FDG:

Fluorodeoxyglucose

i-PET:

Interim-PET

OSEM:

Ordered subset expectation maximization

PET-CT:

Positron emission tomography/computed tomography

PM:

Photomultiplier

PSF:

Point-spread-function

QC:

Q. Clear

RRC:

Resolution recovery coefficient

SD:

Standard deviation

Si:

Silicon

SUV:

Standardized uptake value

References

  1. van der Vos CS, et al. Quantification, improvement, and harmonization of small lesion detection with state-of-the-art PET. Eur J Nucl Med Mol Imaging. 2017;44(Suppl 1):4–16.

    PubMed  PubMed Central  Google Scholar 

  2. Barrington SF, et al. Concordance between four European centres of PET reporting criteria designed for use in multicentre trials in Hodgkin lymphoma. Eur J Nucl Med Mol Imaging. 2010;37(10):1824–33.

    Article  Google Scholar 

  3. Biggi A, et al. International validation study for interim PET in ABVD-treated, advanced-stage hodgkin lymphoma: interpretation criteria and concordance rate among reviewers. J Nucl Med. 2013;54(5):683–90.

    Article  CAS  Google Scholar 

  4. Barrington SF, Kluge R. FDG PET for therapy monitoring in Hodgkin and non-Hodgkin lymphomas. Eur J Nucl Med Mol Imaging. 2017;44(Suppl 1):97–110.

    Article  Google Scholar 

  5. Barrington SF, et al. PET-CT for staging and early response: results from the Response-Adapted Therapy in Advanced Hodgkin Lymphoma study. Blood. 2016;127(12):1531–8.

    Article  CAS  Google Scholar 

  6. Cheson BD, et al. Recommendations for initial evaluation, staging, and response assessment of Hodgkin and non-Hodgkin lymphoma: the Lugano classification. J Clin Oncol. 2014;32(27):3059–68.

    Article  Google Scholar 

  7. Hsu DF, et al. Studies of a next generation silicon-photomultiplier-based time-of-flight PET/CT system. J Nucl Med. 2017;58(9):1511–18.

    Article  CAS  Google Scholar 

  8. Boellaard R, et al. FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0. Eur J Nucl Med Mol Imaging. 2015;42(2):328–54.

    Article  CAS  Google Scholar 

  9. Kaalep A, et al. Feasibility of state of the art PET/CT systems performance harmonisation. Eur J Nucl Med Mol Imaging. 2018;45(8):1344–61.

    Article  Google Scholar 

  10. Aide N, et al. EANM/EARL harmonization strategies in PET quantification: from daily practice to multicentre oncological studies. Eur J Nucl Med Mol Imaging. 2017;44(Suppl 1):17–31.

    Article  CAS  Google Scholar 

  11. Ross S, Clear Q. GE Healthcare. White paper; 2014.

    Google Scholar 

  12. Sadik M, et al. Automated quantification of reference levels in liver and mediastinal blood pool for the Deauville therapy response classification using FDG-PET/CT in Hodgkin and non-Hodgkin lymphomas. Clin Physiol Funct Imaging. 2019;39(1):78–84.

    Article  Google Scholar 

  13. Enilorac B, et al. Does PET reconstruction method affect Deauville score in lymphoma patients? J Nucl Med. 2018;59(7):1049–55.

    Article  CAS  Google Scholar 

  14. Sher A, et al. For avid glucose tumors, the SUV peak is the most reliable parameter for [(18) F]FDG-PET/CT quantification, regardless of acquisition time. EJNMMI Res. 2016;6(1):21.

    Article  Google Scholar 

  15. Munk OL, et al. Point-spread function reconstructed PET images of sub-centimeter lesions are not quantitative. EJNMMI Phys. 2017;4(1):5.

    Article  CAS  Google Scholar 

Download references

Acknowledgements

The authors would like to acknowledge the staff in the Department of Clinical Physiology and Nuclear Medicine, Skåne University Hospital, Malmö and Lund, Sweden, for their daily work in including patients for research. We also thank Anna Åkesson for statistical advice.

Funding

The work was made possible by research grants from the Knut and Alice Wallenberg Foundation, the Swedish Federal Government under ALF agreement and from Region Skåne. The funders of the study were not involved in the study design, data collection, data interpretation, writing of the report, nor in the decision to submit the paper for publication. The authors have no commercial interests.

Author information

Authors and Affiliations

Authors

Contributions

JL contributed to the data generation and analysis and wrote the first version of the manuscript. DM contributed to the design of the study, data generation and analysis and was a major contributor in writing the manuscript. LE contributed to the design of the study and software for data generation. PW contributed to the design of the study and to the data analysis. ET conceived of the study and contributed to the design of the study, data generation and analysis and was a major contributor in writing the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to John Ly.

Ethics declarations

Ethics approval and consent to participate

This study was approved by the Regional Ethical Review Board (#2016/417) and was performed in accordance with the Declaration of Helsinki. All patients provided written informed consent.

Consent for publication

All patients provided written informed consent.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Ly, J., Minarik, D., Edenbrandt, L. et al. The use of a proposed updated EARL harmonization of 18F-FDG PET-CT in patients with lymphoma yields significant differences in Deauville score compared with current EARL recommendations. EJNMMI Res 9, 65 (2019). https://doi.org/10.1186/s13550-019-0536-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13550-019-0536-3

Keywords