Skip to main content

Table 3 Comparison between the SiPM-based handheld gamma camera of this study and the CZT-based CrystalCam (Crystal Photonics), both of which have been integrated with declipse®SPECT

From: Evaluation of camera-based freehand SPECT in preoperative sentinel lymph node mapping for melanoma patients

 

SiPM camera

CrystalCam

Detector material

NaI(Tl)

CZT

Active area shape

square

square

FOV (mm)

63 × 63

40 × 40

Active area (sq. mm)

3906

1600

Number of detector elements

625

256

Detector element pitch (mm)

2.5

2.5

Energy resolution at 140 keV

22%

 < 7%

Overall physical dimensions (mm)

75 (dia) × 41 (H)

60 × 60 × 160 (H)

Mass (kg)

1.4

0.8