Skip to main content

Table 6 The mean activity (kBq/mL) and the percentage standard deviation (%SD) measured from the uniform compartment of the phantom

From: Evaluation of image quality with four positron emitters and three preclinical PET/CT systems

PET/CT system 11C
Mean activity (kBq/mL)
11C
%SD
68Ga
Mean activity (kBq/mL)
68Ga
%SD
18F
Mean activity (kBq/mL)
18F
%SD
89Zr
Mean activity (kBq/mL)
89Zr
%SD
RAYCAN 190.13 5.81 168.90 6.91 182.57 6.03 196.61 6.24
Inveon 171.57 5.99 170.16 7.05 185.50 4.85 205.30 5.69
Molecubes 186.67 8.46 143.94 7.16 174.46 7.39 154.20 7.34