Table 1 Spatial resolutions measured on the reconstructed images obtained with the 85 keV ± 20% energy window and on the sum of the images obtained with each energy window (85 keV ± 20%, 154 keV ± 10%, 270 keV ± 10%)

From: 223Ra-dichloride therapy of bone metastasis: optimization of SPECT images for quantification

View

FWHM (mm) without attenuation

FWHM (mm) with attenuation

85 keV

Sum

Difference (%)

85 keV

Sum

Difference (%)

Axial

15.8

16.3

3.6

17.2

17.1

− 1.0

Sagittal

16.3

16.0

− 1.5

17.6

16.5

− 6.9