Fig. 7From: Performance evaluation of the next generation solid-state digital photon counting PET/CT systemTiming resolution and energy resolution of the dPET system. a NEMA NU2-2018 timing resolution in FWHM and FWTM as a function of activity concentration. b Stability of timing resolution (22Na point source based) within the 31-month monitoring time window. It demonstrates a slight improvement from 325Â ps to 322Â ps when system calibration was performed 6Â months after installation. c Stability of energy resolution during the monitoring time periodBack to article page