Skip to main content

Table 2 Recovered activity in the tumor insert before and after compensation for partial volume effect through recovery coefficients. Recovered activity concentrations are compared to measured values

From: Phantom validation of quantitative Y-90 PET/CT-based dosimetry in liver radioembolization

Day of scan

True activity concentration (MBq mL−1)

Average recovered concentration (MBq mL−1) [1st–3rd quartile] (no recovery coefficient applied)

Average recovered concentration (MBq mL−1) [% deviation] (recovery coefficient applied)

Homogeneity Index

Reconstruction parameters

Scan Time (min)

1

5.50 ± 0.16

4.30 [3.24–5.36]

5.3 [− 3.6%]

0.34

2i, 15s

15

4

2.52 ± 0.08

1.95 [1.4–2.51]

2.41 [− 4.4%]

0.42

2i, 15s

15

5

1.94 ± 0.06

1.51 [1.12–1.88]

1.86 [− 4.1%]

0.36

2i, 15s

30

6

1.49 ± 0.04

1.03 [0.72–1.34]

1.27 [− 14.7%]

0.42

2i, 15s

30

12

0.31 ± 0.01

0.22 [0.076–0.294]

0.27 [− 12.4%]

1.01

2i, 15s

30