From: Initial evaluation of the Celesteion large-bore PET/CT scanner in accordance with the NEMA NU2-2012 standard and the Japanese guideline for oncology FDG PET/CT data acquisition protocol version 2.0
Sphere diameter (mm)
Contrast recovery
Background variability
10
27.9
6.2
13
48.6
5.3
17
52.0
4.6
22
60.5
4.1
28
72.9
3.9
37
77.1
3.8