Table 1 Unit operations and select parameters

From: Simplified programming and control of automated radiosynthesizers through unit operations

Unit operation

Definition

Parameter

Description

ADD

Adds a reagent to a reaction vessel

Reagent

Reagent to add

Reactor

Reactor to which the reagent will be added

Delivery position

Choice of two separate delivery positions

EVAPORATE

Evaporates the contents of a reaction vessel

Reactor

Reactor to evaporate

Evaporation temp

Temperature to heat to during evaporation (°C)

Evaporation pressure

Inert gas pressure to use when evaporating (psi)

Final temp

Temperature to cool to after evaporation (°C)

Duration

Time to evaporate after reaching temp (s)

Stir speed

Rate of stirring during evaporation

TRANSFER

Transfers the contents of one reactor to the next (optionally through a purification cartridge)

Source reactor

Reactor to transfer from

Target reactor

Reactor to transfer to

Mode

‘Trap’ to send to waste or ‘Elute’ to send to the next reactor

REACT

Seals the reactor and heats

Reactor

Reactor to heat

Reaction temp

Temperature to heat to during reaction (°C)

Reaction position

Choice of two separate reaction positions to avoid cross-contamination if multiple sealed reactions are performed in the same vessel

Duration

Time to react after reaching temperature (s)

Final temp

Temperature to cool to after reaction (°C)

Stir speed

Rate of stirring during reaction

PROMPT

Pauses the sequence run and prompts the user

Message

Message to display when prompting the user

INSTALL

Moves the reactor to the install position for reaction vessel removal and/or installation and prompts the user

Reactor

Reactor to move to the install position

Message

Message to display when prompting the user

COMMENT

User comment

Comment

User-specified comment for documentation purposes only (no action performed)

TRAPF18

Traps [18F]fluoride on a QMA cartridge

Reactor

Reactor where the QMA cartridge is located

Cyclotron flag

Specifies if the cyclotron or ELIXYS will push the solution

Duration

Trap time (s)

Pressure

Pressure of inert gas to use when trapping (psi)

ELUTEF18

Uses a reagent to elute [18F]fluoride off a QMA cartridge

Reactor

Reactor where the QMA cartridge is located

Reagent

Reagent with which to elute the contents of the QMA cartridge

Duration

Elute time (s)

Pressure

Pressure of inert gas to use when eluting (psi)

MIX

Mixes the contents of a reactor by stirring

Reactor

Reactor to mix

Duration

Mix time (s)

Stir speed

Rate of stirring while mixing

EXTERNALADD

Allows the user to externally add a reagent via tubing

Reactor

Reactor to externally add the reagent

Reagent name

Name of the reagent to add

Message

Message to display when prompting the user

TRANSFERTOHPLCa

Transfers the contents of the reactor to the HPLC injection loop

Source reactor

Reactor to transfer from

Mobile phase

Reagent containing the mobile phase

MEASURERADIATIONa

Measures the radiation levels

Reactor

Reactor to measure

  1. QMA quaternary methylammonium. aThese unit operations are currently under development.