Skip to main content

Table 3 Preclinical PET motion effect measurements

From: Gating, enhanced gating, and beyond: information utilization strategies for motion management, applied to preclinical PET

 

Units

Measurements

Gated

Gating+

Gating+ with phase shift

Ungated

Number of frames (phase increment = 360°/number)

   

16

16

90

1

Kidney VOI, n = 102

Average kidney uptake

Average VOI value, average of all gates

Relative values

1.01

1.01

1.01

1.00

% SD

Average uptake across gates

Percentage

2.69

1.47

1.47

 

Maximum displacement

Average of all scans (SD of all scans)

mm

1.50 (0.56)

1.10 (0.75)

1.10 (0.75)

 

Liver profile, n = 34

Liver boundary displacement

 

mm

3.25

3.04

3.05

 

Shoulder VOI, n = 84

% SD in VOI

Average of all gates

Relative values

1.53

1.05

1.05

1.00

Global COM, n = 84

Maximum COM displacement

Average of all scans (SD of all scans)

mm

0.71 (0.46)

0.43 (0.41)

0.43 (0.41)

 
  1. Summary of VOI, line profile, and COM measurements in ungated, gated, and gating+ images across the population of preclinical PET scans. All numbers are average values of population or all appropriate population scans.