Skip to main content

Table 3 %SDu and the relative error between average activity measured and injected activity

From: dAcquisition setting optimization and quantitative imaging for 124I studies with the Inveon microPET-CT system

Isotope

Correction

%SDu

Error (%)

  

2DFBP

2DOSEM

3DOSEM/MAP

2DFBP

2DOSEM

3DOSEM/MAP

18F

No correction

6.3

6.7

6.2

-28.3

-28.4

-25.2

 

AC

5.0

5.5

5.8

0.6

0.5

6.2

 

AC-SC

5.1

5.7

4.8

-1.7

-2.1

3.7

124I

No correction

14.3

16.6

15.6

-36.2

-34.7

-34.3

 

AC

13.8

16.3

14.5

-8.0

-6.9

-3.6

 

AC-SC

14.4

17.8

15.2

-13.7

-13.4

-9.9

 

AC-SC-GC

16.9

21.3

14.2

-27.2

-26.6

-23.0

  1. %SDU, percentage standard deviation; 2DFBP, 2D filtered back-projection; 2DOSEM, 2D ordered-subset expectation maximization; 3DOSEM/MAP, 3DOSEM in combination with maximum a posteriori; AC, attenuation correction; SC, scatter correction; GC, gamma-coincidence correction.