From: Performance and application of the total-body PET/CT scanner: a literature review
Metrics | PennPET explorer | Quadra | uEXPLORER |
---|---|---|---|
aFOV | 64Â cm (prototype) 112Â cm (upgraded) | 106Â cm | 194Â cm |
Sensitivity (3D) | 54 kcps/MBq | 83.4 kcps/MBq (MRD 85) 176.0 kcps/MBq (MRD 322) | 171 kcps/MBq |
Space resolution@1Â cm* | 4.0Â mm | 3.8Â mm | 2.9Â mm |
Space resolution@10 cm* | – | – | 3.2 mm |
scatter fraction | 32% | 36% (MRD 85) 37% (MRD 322) | 38% |
P-NECR | 1550 kcps@25Â kBq/cc | 1613 kcps@27.49Â kBq/ml (MRD 85) 2956 kcps@27.49Â kBq/ml (MRD 322) | 1549 kcps@16.8Â kBq/cc |
Time resolution | 256Â ps | 228Â ps (MRD 85) 230Â ps (MRD 322) | 430Â ps |
Energy window range (3D) | 450–630 keV | 435–585 keV | 430–650 keV |
Data collection method | 3D, List-mode, Static, Dynamic | 3D, List-mode, Static, Dynamic | 3D, List-mode, Static, Dynamic, Gating |