Quantitation of mitral regurgitation using positron emission tomography

Abstract

Background

Cardiac positron emission tomography (PET) offers non-invasive assessment of perfusion and left ventricular (LV) function from a single dynamic scan. However, no prior assessment of mitral regurgitation severity by PET has been presented. Application of indicator dilution techniques and gated image analyses to PET data enables calculation of forward stroke volume and total LV stroke volume. We aimed to evaluate a combination of these methods for measurement of regurgitant volume (RegVol) and fraction (RegF) using dynamic 15O-water and 11C-acetate PET in comparison to cardiovascular magnetic resonance (CMR).

Results

Twenty-one patients with severe primary mitral valve regurgitation underwent same-day dynamic PET examinations (15O-water and 11C-acetate) and CMR. PET data were reconstructed into dynamic series with short time frames during the first pass, gated 15O-water blood pool images, and gated 11C-acetate myocardial uptake images. PET-based RegVol and RegF correlated strongly with CMR (RegVol: 15O-water r = 0.94, 11C-acetate r = 0.91 and RegF: 15O-water r = 0.88, 11C-acetate r = 0.84, p < 0.001). A systematic underestimation (bias) was found for PET (RegVol: 15O-water − 11 ± 13 mL, p = 0.002, 11C-acetate − 28 ± 16 mL, p < 0.001 and RegF: 15O-water − 4 ± 6%, p = 0.01, 11C-acetate − 10 ± 7%, p < 0.001). PET measurements in patients were compared to healthy volunteers (n = 18). Mean RegVol and RegF was significantly lower in healthy volunteers compared to patients for both tracers. The accuracy of diagnosing moderately elevated regurgitant volume (> 30mL) was 95% for 15O-water and 92% for 11C-acetate.

Conclusions

LV regurgitation severity quantified using cardiac PET correlated with CMR and showed high accuracy for discriminating patients from healthy volunteers.

Introduction

Primary mitral valve regurgitation is a common valve disease with increasing prevalence, primarily affecting the elderly [1,2,3]. Mitral regurgitation leads to left ventricular (LV) remodeling and eventually dysfunction. The treatment options are surgical valve repair, replacement or percutaneous intervention. Correct timing of treatment is a complex task that requires monitoring with cardiac imaging. In routine practice, echocardiography is the preferred method, and image findings are combined with clinical assessment [4, 5]. For evaluating the severity of mitral regurgitation, several qualitative signs as well as measurement of regurgitant volume (RegVol) and regurgitant fraction (RegF) are performed. LV volumes, and LV ejection fraction (LVEF) are recommended for guiding decisions. The gold standard for these assessments is cardiovascular magnetic resonance (CMR) [6].

Recent evidence suggests that besides the standard measurements, additional assessment of myocardial performance is important [7]. Signs of myocardial damage and remodeling is likely an earlier indication of decompensated valvular disease [8,9,10]. In light of this, despite of its limited use today, molecular imaging with positron emission tomography (PET) could be of increased relevance in mitral regurgitation. In a recent study, our group evaluated myocardial external efficiency as an outcome variable in severe mitral regurgitation using 11C-acetate PET [11]. PET added incremental information to CMR measurements of regurgitant magnitude, and was a significant predictor of outcome on its own. A dual-modality approach of CMR and 11C-acetate PET would be a powerful tool, but resource demanding. It is therefore of interest to investigate the use of stand-alone PET to assess regurgitation severity.

Calculation of the parameters needed for such evaluations has previously been studied using dynamic 11C-acetate imaging. First-pass analysis and indicator dilution techniques enable quantification of cardiac output (CO) and forward stroke volume (FSV) [12, 13]. Electrocardiography (ECG)-gating of 11C-acetate late uptake images can provide LV total stroke volume (LVSV) from the same acquisition [14]. Hence, using a combination of indicator dilution techniques and ECG-gating, it should be possible to estimate RegVol and RegF from one dynamic 11C-acetate PET scan.

As cardiac PET is mainly utilized to evaluate ischemic heart disease, it would likely be further beneficial to use clinically important perfusion tracers for this purpose. 15O-water is considered the reference method for quantifying myocardial blood flow (MBF) [15,16,17]. CO and FSV can be calculated from dynamic 15O-water PET using the same methods as presented for 11C-acetate [13]. ECG-gating of 15O-water for assessment of LVSV constitutes a challenge since water is freely diffusible and yields no retention images. However, methods for blood pool-based gating are feasible [18, 19]. As such, RegVol and RegF measurements should also be attainable with 15O-water PET.

In this study, we aimed to investigate using both a retention PET tracer (11C-acetate) and a freely diffusible PET tracer (15O-water), for quantitation of RegVol and RegF, as compared to CMR.

Methods

Study design and population

The patient data used in the current study have been described in detail previously [11]. We utilized a sub-cohort of this previous study, including only subjects where ECG-gated 15O-water data were available. All patients were diagnosed with asymptomatic severe degenerative and chronic mitral regurgitation, as according to echocardiographic criteria. Subjects with more than mild concomitant valve diseases were excluded. Patients underwent 15O-water PET, 11C-acetate PET and CMR examinations on the same day. PET and CMR scans were conducted within one hour, avoiding fluid intake between scans to minimize changes in hemodynamic loading conditions.

In addition, 21 healthy volunteers were included and underwent 15O-water PET. Five of these additionally underwent 11C-acetate PET and echocardiography on the same day. Healthy volunteers had no history of cardiovascular disease or symptoms at the time of the examinations.

Imaging data were acquired in three studies, all approved by the Regional Ethical Review Board at Uppsala University (Dnr: 2012 − 543, 2020–07017, 2021–05230). The examinations were performed at Uppsala University Hospital between 2016 and 2023 and all subjects provided written informed consent.

PET scanning procedure

PET examinations were conducted with a digital PET/CT scanner (Discovery MI, GE Healthcare) with an axial detector coverage of 20 cm (DMI-20: all patients, 5 healthy volunteers with 15O-water and 11C-acetate scans) or 25 cm (DMI-25: 16 healthy volunteers with 15O-water scans). The scans were performed in the resting state and a 4-lead ECG was connected during the procedure. Manual heart rate recording was performed before the start of the scans and during first pass.

A low dose CT was acquired for attenuation correction and anatomical localization. Subsequently, 400 MBq of 15O-water was injected in an antecubital vein using an automated bolus (10 mL at 0.8 mL/s followed by 30 mL saline at 2.0 mL/s) simultaneous with the start of a 4 min PET-acquisition in list-mode. After approximately 15 min post the 15O-water injection, 5 MBq/kg 11C-acetate was administered using the same injection protocol and a 27 min list-mode acquisition.

List-mode PET data were reconstructed into dynamic series (15O-water: 1 × 10, 8 × 5, 4 × 10, 2 × 15, 3 × 20, 2 × 30s and 11C-acetate: 12 × 5, 6 × 10, 4 × 30, 4 × 60, 2 × 120, 3 × 300s) and ECG-gated images with 16 bins. The 15O-water gated images were reconstructed to represent the blood pool using a standardized and fixed time window of 0–50 s. Since no previous studies have assessed blood pool 15O-water gating using more than 8 bins, an additional 8-bin reconstruction was performed for comparison. The 11C-acetate gated reconstructions represented the myocardial retention 2–7 min post injection. PET images were reconstructed using time-of-flight ordered subsets expectations maximization (3 iterations, 16 subsets, 5 mm Gaussian post-filter) with point spread function recovery.

PET image analysis

All PET images were analysed using aQuant Research (MedTrace A/S, Hørsholm, Denmark). The 15O-water gated analysis was based on blood-pool images and the steepest path approach, defining seed points in the image, as described previously [20, 21]. Parametric blood-volume images were derived from the single-tissue compartment model for 15O-water [22], allowing for automatic separation of the seeds in the left and right side of the heart. The gated analysis included only the LV. The LV was separated from the left atrium in each bin using a line placed in the valve plane, and the seeds belonging to the LV-side were used for volume calculation. Manual adjustments of the separation between left atrium and LV were performed when deemed necessary by the observer, blinded to CMR data. A representative 15O-water gating image is shown in Fig. 1.

Fig. 1
figure 1

Example of blood-pool based gating in 15O-water PET during the end-diastolic (A) and end-systolic (B) phase

The 11C-acetate gated analysis was fully automated and based on LV wall delineations [14], applied to all binned images. The gated analyses were used to measure LV end-diastolic volume (EDV) and end-systolic volume (ESV). Total LV stroke volume (LVSV) was defined as EDV-ESV, and LV ejection fraction (LVEF) as LVSV/EDV. An animated example of 15O-water and 11C-acetate gating in the same patient is shown Supplementary material (Figure S1).

Cluster analysis was used to extract both the arterial input function consisting of the LV and aorta, and the venous input function consisting of the right atrium, right ventricle and pulmonary trunk. Cardiac output (CO) was calculated using indicator dilution methods based on the area under the curve (AUC) of the arterial and venous input, with the formula CO = Injected tracer dose/AUC [12, 13]. Heart rate (HR) registered during the first pass was used to calculate FSV, defined as CO/HR. HR in 15O-water scans was derived from the ECG-gated data, as this resembled the actual first pass (0–50 s), whilst a manually recorded HR was used for the 11C-acetate scans. A calibration of the PET FSV was performed using linear regression of the 11C-acetate FSV towards CMR. The regression was subsequently applied prospectively to 15O-water imaging data. PET-based regurgitant volume (RegVol) was estimated by subtraction of calibrated FSV from LVSV, and subsequently, regurgitant fraction (RegF) was calculated as RegVol/LVSV.

Cardiovascular magnetic resonance (CMR)

The CMR examinations were performed on a 3-Tesla Ingenia Philips whole body scanner (Philips Healthcare, Best, Netherlands) with an 80 mT*m− 1 gradient system. Short- and long-axis cine images were generated with a steady-state free precession pulse sequence. The basal slice was defined on long-axis images and LV geometrical volumes were manually segmented from short-axis images. End-diastolic endo- and epicardial contours were delineated, including the papillary muscle tissue in the blood pool, as according to clinical procedure. Additionally, LV volume calculations were repeated with manual exclusion of the papillary muscles from the blood pool. Aortic FSV was quantified semiautomatically with phase-contrast images acquired perpendicular to the proximal ascending aorta during free breathing. Aortic FSV was corrected for aortic backflow, to better resemble the PET-calculations. Images were analysed using commercial software (CVI42, Circle Cardiovascular Imaging, Calgary, Canada). RegVol and RegF were calculated as described in the PET image analysis section.

Statistics

Continuous variables were tested for normal distribution using Shapiro-Wilks test and were normal distributed if not stated otherwise. Continuous variables are expressed as mean ± standard deviation (SD). Correlations between measurements were assessed with linear regression and agreement between methods with Bland-Altman analysis. Systematic and proportional bias were tested with paired t-tests and linear regression, respectively. Unpaired t-tests were used to compare PET measurements in patients and healthy volunteers. Diagnostic accuracy was evaluated with contingency tables. P-values < 0.05 were considered statistically significant. Statistical and graphical analyses were performed in JMP 17 (SAS Institute Inc., Cary, NC, USA) and GraphPad Prism version 10 (GraphPad Software, La Jolla, CA, USA).

Results

Patient data are presented in Table 1. In total, 21 patients were included. All patients underwent 15O-water PET and CMR scanning, and 20 underwent 11C-acetate PET. One patient did not undergo 15O-water gated analysis due to technical issues and one other patient did not complete the cine CMR sequence for volumetric assessment. Thus, FSV was obtained with CMR and 15O-water in all 21 patients, and LVSV, RegVol and RegF in 20 patients. The extra CMR analysis with exclusion of papillary muscle from the blood pool was performed in 19 patients.

Table 1 Clinical data of patients with primary mitral valve regurgitation and healthy volunteers. Asterisk indicates statistical significance in unpaired t-test

Of the 20 patients with CMR measurements of RegVol, 16 were confirmed to have severe mitral regurgitation (> 60 mL), 3 patients were found to have moderate disease (> 30 < 60 mL) and 1 mild disease (< 30 mL).

Two healthy volunteers with imaging findings suggestive of undiagnosed cardiac disease (significant chamber dilatation, LV hypertrophy) were excluded. One healthy volunteer was excluded due to excessive motion during the PET acquisition, compromising the analysis. Ultimately, the total number of healthy volunteers included in the analysis were 18 for 15O-water, of which 5 also underwent 11C-acetate PET.

Gated analysis and volumetric measurements

PET LV volumes correlated well with CMR for both 15O-water and 11C-acetate (EDV r = 0.97 and 0.95; ESV r = 0.75 and 0.88; LVSV r = 0.94 and 0.92, all p < 0.001) and moderately for LVEF (r = 0.46, p = 0.05 and r = 0.54, p = 0.02). PET underestimated LV volumes, while no systematic bias was present for LVEF when calculated with 16 bins. Mean 15O-water and 11C-acetate values from the gated image analyses, and results from Bland-Altman analyses of EDV, ESV, LVSV, and LVEF compared to CMR are presented in Table 2.

Table 2 Mean values ± standard deviation (SD) and comparison of values from volumetric CMR measurements and PET gating analysis in the mitral regurgitation patients. Bias was calculated using bland-Altman analysis and p-values were calculated with paired t-tests. EDV = end-diastolic volume, ESV = end-systolic volume, SV = total stroke volume, EF = ejection fraction

The 15O-water LVSV calculated with 16 bin gating showed slightly stronger correlation and closer agreement towards CMR, compared to 8 bins (r = 0.94 vs. 0.92, bias=-11.6 ± 15.0 mL vs. -26.3 ± 17.8 mL). Hence, the 16-bin reconstruction was used to calculate RegVol and RegF in 15O-water PET. A comparison between 16 bin 15O-water, 11C-acetate and CMR based LVSV is shown in Fig. 2.

Fig. 2
figure 2

Scatter plots (left) and Bland-Altman plots (right) of left ventricular total stroke volume (LVSV) calculated with 15O-water (WAT), 11C-acetate (ACE) and cardiovascular magnetic resonance (CMR). Dashed lines are lines of identity (A, C, E), dotted lines limits of agreement (B, D, F), and solid lines represent linear regression and mean bias (B, D, F)

First-pass analysis with indicator dilution techniques

PET-based CO calculated from the arterial and venous input function, respectively, correlated strongly (15O-water r = 0.94 and 11C-acetate r = 0.97, p < 0.001) (Figure S2), and a similar result was found for FSV (15O-water: r = 0.94 and 11C-acetate r = 0.97, p < 0.001). For each tracer, no systematic or proportional bias was found between arterial and venous input function.

Uncalibrated PET-based FSV was significantly higher than CMR, when calculated from both arterial and venous clusters (15O-water arterial bias = 35.7 ± 9.7 mL, venous bias = 36.7 ± 9.4 mL and 11C-acetate arterial bias = 37.7 ± 12.6 mL, venous bias 35.7 ± 12.6 mL, p < 0.001). The 11C-acetate venous input function was used to calibrate all PET FSV values. Linear regression with uncalibrated FSV versus CMR resulted in an intercept of 9 and a slope of 0.59, which was applied as a calibration factor for the arterial and venous cluster in both tracers. Linear regression analyses for uncalibrated and calibrated FSV values are shown in Supplementary material (Figure S3). The calibrated FSV values from the arterial input for 15O-water and 11C-acetate were used to calculate PET RegVol and RegF.

Calibrated FSV in 15O-water and 11C-acetate correlated strongly with CMR. No systematic bias was found, but a small significant proportional bias for 15O-water versus both CMR and 11C-acetate (Fig. 3).

Fig. 3
figure 3

Scatter plots and Bland-Altman plots of forward stroke volume (FSV) calculated with 15O-water (WAT), 11C-acetate (ACE) and cardiovascular magnetic resonance (CMR). Dashed lines are lines of identity (A, C, E), dotted lines are limits of agreement (B, D, F), and solid lines represent linear regression and mean bias (B, D, F). A proportional bias between 15O-water versus CMR (B) and 15O-water versus 11C-acetate (F) is present

Regurgitant volume and fraction

Strong correlations were found for RegVol and RegF based on 15O-water and 11C-acetate PET versus CMR. For both PET tracers, RegVol and RegF were underestimated. There was a closer agreement between 15O-water and CMR versus 11C-acetate and CMR, and a significant systematic bias was present between the tracers (Figs. 4 and 5). There was a proportional bias when comparing 15O-water and 11C-acetate RegF (slope = 0.23, p = 0.01).

Fig. 4
figure 4

Scatter plots and Bland-Altman plots of regurgitant volume (RegVol) calculated with 15O-water, 11C-acetate and cardiovascular magnetic resonance (CMR). Dashed lines are lines of identity (A, C, E), dotted lines are limits of agreement (B, D, F), and solid lines represent linear regression and mean bias (B, D, F)

Fig. 5
figure 5

Scatter plots and Bland-Altman plots of regurgitant fraction (RegF) calculated with 15O-water (WAT), 11C-acetate (ACE) and cardiovascular magnetic resonance (CMR). Dashed lines are lines of identity (A, C, E), dotted lines are limits of agreement (B, D, F), and solid lines represent linear regression and mean bias (B, D, F). A proportional bias between 11C-acetate versus 15O-water (F) is present

Mean values and comparisons between PET-parameters in patients and healthy volunteers are presented in Table 3. Average PET RegVol and RegF for both tracers were significantly lower in healthy volunteers compared to the patients (RegVol: 15O-water = 21 ± 7 vs. 78 ± 39 mL, 11C-acetate = 4 ± 8 vs. 62 ± 34 mL and RegF 15O-water = 22 ± 7 vs. 48 ± 13%, 11C-acetate = 5 ± 10 vs. 42 ± 13%, p < 0.001). Contingency tables showed that moderate mitral regurgitation (RegVol > 30 mL) was detected with an accuracy of 95% (n = 38, p < 0.001) for 15O-water and 92% (n = 25, p < 0.001) for 11C-acetate.

Table 3 Results from PET-measurements comparing mitral regurgitation patients and healthy volunteers: mean values ± standard deviation (SD) and two-tailed p-values from t-test between patients and healthy volunteers. MR = mitral regurgitation, HV = healthy volunteer, LV = left ventricular

When the LV papillary muscle volume was excluded from the CMR LVSV measurements, no significant bias was found between 15O-water PET and CMR for RegVol and RegF (RegVol bias=-2.7 ± 13.9 mL, p = 0.4 and RegF bias=-0.9 ± 5.4%, p = 0.5). However, a significant bias remained for 11C-acetate PET (RegVol bias=-21.0 ± 17.2 mL, p < 0.001 and RegF bias=-7.6 ± 7.5%, p = 0.001).

Both 8 and 16 bin based 15O-water LVSV correlated strongly with CMR for RegVol and RegF calculation. Using 16 bins lead to stronger correlations and higher precision versus 8 bins (RegVol: r = 0.94 vs. 0.90, bias=-11.0 ± 13.4 mL vs. -25.8 ± 17.1 mL and RegF: r = 0.88 vs. 0.82, bias=-4.0 ± 6.2% vs. 9.5 ± 8.2%).

Discussion

Several studies evaluated PET-based LV volumes and cardiac function with ECG gating and indicator dilution techniques, but, to the best of our knowledge, this is the first study combining the two methods for quantitation of mitral regurgitation severity. The different functional assessments required for regurgitation calculation were possible to perform with a high degree of automation on data from a single PET acquisition, using either a freely diffusible (15O-water) or a metabolically trapped (11C-acetate) PET tracer with similar results. PET measurements of regurgitation magnitude correlated strongly with CMR, and using PET-based RegVol it was possible to discriminate the controls from patients with high accuracy. These results indicate that PET is able to detect and roughly quantify mitral regurgitation.

All gated PET reconstructions used in the current study can be performed automatically on the PET console, if the time window for ECG gating is pre-defined. For 11C-acetate, this is straightforward, as the myocardial uptake is used to delineate the cavity and calculate volumes. We chose 11C-acetate data between 2 and 7 min, as has been described previously [14]. During this time window, the myocardial uptake is expected to be clearly visualized, but a number of different time settings would likely be applicable. 15O-water blood pool gating is more sensitive, since for adequate contrast between myocardial tissue and cavity, the acquisition window has to be short enough for the bolus to remain intact during first pass, which limits the amount of counts available for calculation. The acquisition window might need adjustments for patients with unusually low or high CO [18]. A window of 0–50 s worked for the analysis of all the subjects included in the current study, and CO did not differ between patients and healthy volunteers. If the scans are acquired in list-mode, as in this study, it would be possible to reconstruct the gated series using a different acquisition interval retrospectively.

PET-based LV volumes correlated strongly with CMR, while the correlation for LVEF was only moderate. This was most likely due to the narrow range of LVEF values, and in good agreement with previously reported measurements using ECG gated PET [9]. The accuracy of gating-based volume measurements is dependent on temporal resolution and no published data have been presented where more than 8 gating bins were used for 15O-water PET, likely due to the use of data from older, non-digital PET scanners with relatively low sensitivity and resolution. Increasing the number of bins to 16 improved the agreement of PET-based RegVol and RegF towards CMR. A slightly higher change was seen in ESV as compared with EDV when increasing the number of gating bins. This was expected since the end-systolic phase is shorter, and the ESV is more likely overestimated when using a low temporal resolution.

PET systematically underestimated LV volumes in comparison to CMR for both tracers. However, when performing an additional CMR analysis with removal of papillary muscle volume from EDV, the bias was no longer present between 15O-water and CMR. Also, the underestimation for RegVol and RegF was eliminated. In hindsight, this observation suggests that CMR-based evaluation of LV regurgitation using standard circular chamber delineations overestimates RegVol. The bias for LVSV, RegVol and RegF remained between 11C-acetate and CMR, primarily explained by the larger underestimation of EDV in 11C-acetate. This underestimation for 11C-acetate EDV was similar to prior results [14].

The 15O-water and 11C-acetate arterial and venous based CO correlated strongly, with no bias between tracers or clusters. This was in line with the high accuracy presented in previous studies assessing CO from PET data with older scanner types [12, 13], and suggests that the method is robust and reproducible. With a recorded HR, the FSV is thereafter easily derived from the CO. Using ECG during the scan automated this process for 15O-water PET, since the HR registration corresponded to the first pass during which CO was measured. For 11C-acetate PET, the ECG-derived HR corresponded to 2–7 min into the scan, and thus it is possible that the HR recorded differs from the first pass. This was the case in 2 subjects, where the gating-based and manual HR differed 10 and 13 heart beats. Therefore, the manual recording was used for calculation of 11C-acetate FSV, even though there were no significant mean differences between the two methods on the group level.

As in all cases of indicator dilution techniques, correct PET-based CO calculation requires the knowledge of exact amount of injected radioactivity. Since 15O-water has a half-life of 2.03 min, precise measurements are challenging. The scans conducted in the current study utilized a standardized semi-automatic tracer injection. Radioactivity was measured in a syringe in a dose calibrator 2 min before injection, aiming for 830 MBq to account for isotope half-life and residual activity in the syringe and tubing, and resulting in an injected dose of approximately 400 MBq. It is difficult to obtain the precise amount for every injection and the residue depends somewhat on the concentration of the tracer produced, and the manual handling of the syringe. Using more automated boluses would likely ensure an even more robust CO calculation for 15O-water. With 11C-acetate, the residual activity in the syringe and tubing could be accounted for by direct measurements after injection. However, the actual amount of radioactivity reaching the main circulation during the first pass is still somewhat uncertain as PET images frequently show remaining activity at the injection site and brachial veins, even when flushing with 30 mL saline. This may contribute to the overestimation of FSV as shown in Figure S3.

In line with the findings in this study, earlier experiments have shown PET-based FSV utilizing 5 s frames during first pass to be overestimated in comparison to CMR [13]. The overestimation was scanner dependent, but no evaluation has previously been performed for the type of scanner used in this study. Therefore, we performed the scanner specific calibration of the FSV towards CMR using linear regression. It is a methodological limitation that cardiac output and FSV calculations currently require scanner-dependent corrections. The calibration should preferably be based on invasive thermodilution or local CMR devices and analysis tools.

In order to utilize either 15O-water or 11C-acetate, an on-site cyclotron is required, which is a barrier for widespread clinical implementation. However, as shown here with 15O-water PET, a technique based on first-pass analysis might allow most PET tracers to be used, provided that dynamic scanning is acquired, reliable estimates of injected dose are available, and the radiopharmaceutical is delivered as a fast and standardized bolus. Our results indicate that PET could be used for simultaneous evaluation of primary or secondary mitral regurgitation, LV dilatation and PET-specific parameters such as metabolism or ischemia. This might lead to a more comprehensive diagnosis from cardiac PET scans, while possibly speeding up the patient management. Considering the increasing burden of valvular diseases, allowing for calculation of mitral regurgitation using PET could be valuable. PET-based measurements of regurgitation could be particularly useful to discover secondary mitral regurgitation in perfusion imaging on rare occasions when echocardiography has not been performed prior to the PET-examination.

Study limitations

Some limitations of the present study should be noted. The accuracy of quantifying low levels of RegVol and RegF was not assessed as the healthy volunteers were not examined with CMR, and few patients with mild and moderate mitral regurgitation were included. Only five of the healthy volunteers performed echocardiography, and thus for 13 of the controls included it is unknown whether any of the volunteers had symptom-free, undiagnosed valve disease that could have affected the results. The 15O-water based mean RegVol and RegF was relatively high and likely overestimated in the healthy volunteers (21 mL and 22% respectively). One methodological explanation is that identical settings were used in the gating analysis for patients and controls, although LV volumes are expected to be significantly higher in patients. A further explanation is that the FSV calibration was only performed on data from patients, as only they underwent both CMR and PET. These findings indicate that further development of the methods is needed, incorporating larger cohorts with variation in cardiac function and size. However, the PET measurements were able to separate with confidence patients with substantial mitral regurgitation from normal controls.

An inherent limitation of the PET method is the current inability to distinguish between aortic and mitral regurgitation, since only the total amount of blood not moving forward in the system is taken into account when calculating RegVol and RegF. This suggests that elevated levels of left ventricular regurgitation found with PET should be complemented with further cardiac imaging.

Conclusions

Left ventricular regurgitation can be quantified from a cardiac PET examination. Regurgitant volumes and fractions calculated using two radiopharmaceuticals with different approaches to geometric assessments were reproducible and correlated strongly with gold standard CMR. Using PET-based regurgitant volume, it was possible to discriminate healthy controls from patients with high accuracy.

Data availability

The data to support the findings in the study are available from the corresponding author upon reasonable request.

References

  1. Nkomo VT, Gardin JM, Skelton TN, Gottdiener JS, Scott CG, Enriquez-Sarano M. Burden of valvular heart diseases: a population-based study. Lancet. 2006;368(9540):1005–11.

    Article  PubMed  Google Scholar 

  2. Andell P, Li X, Martinsson A, Andersson C, Stagmo M, Zoller B, et al. Epidemiology of valvular heart disease in a Swedish nationwide hospital-based register study. Heart. 2017;103(21):1696–703.

    Article  PubMed  Google Scholar 

  3. Dziadzko V, Clavel MA, Dziadzko M, Medina-Inojosa JR, Michelena H, Maalouf J, et al. Outcome and undertreatment of mitral regurgitation: a community cohort study. Lancet. 2018;391(10124):960–9.

    Article  PubMed  PubMed Central  Google Scholar 

  4. Otto CM, Nishimura RA, Bonow RO, Carabello BA, Erwin JP 3rd, Gentile F, et al. 2020 ACC/AHA Guideline for the management of patients with Valvular Heart Disease: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice guidelines. Circulation. 2021;143(5):e72–227.

    PubMed  Google Scholar 

  5. Vahanian A, Beyersdorf F, Praz F, Milojevic M, Baldus S, Bauersachs J, et al. 2021 ESC/EACTS guidelines for the management of valvular heart disease. Eur Heart J. 2022;43(7):561–632.

    Article  PubMed  Google Scholar 

  6. Myerson SG, d’Arcy J, Christiansen JP, Dobson LE, Mohiaddin R, Francis JM, et al. Determination of clinical outcome in Mitral Regurgitation with Cardiovascular magnetic resonance quantification. Circulation. 2016;133(23):2287–96.

    Article  PubMed  Google Scholar 

  7. Ajmone Marsan N, Delgado V, Shah DJ, Pellikka P, Bax JJ, Treibel T, et al. Valvular heart disease: shifting the focus to the myocardium. Eur Heart J. 2023;44(1):28–40.

    Article  PubMed  Google Scholar 

  8. Fragasso G, Sanvito F, Monaca G, Ardizzone V, De Bonis M, Pappalardo F, et al. Myocardial fibrosis in asymptomatic patients undergoing surgery for mitral and aortic valve regurgitation. J Cardiovasc Med (Hagerstown). 2022;23(8):505–12.

    Article  CAS  PubMed  Google Scholar 

  9. Ahmed MI, Andrikopoulou E, Zheng J, Ulasova E, Pat B, Kelley EE, et al. Interstitial Collagen Loss, myocardial remodeling, and function in primary mitral regurgitation. JACC Basic Transl Sci. 2022;7(10):973–81.

    Article  PubMed  PubMed Central  Google Scholar 

  10. Miller MA, Adams DH, Pandis D, Robson PM, Pawale A, Pyzik R, et al. Hybrid Positron Emission Tomography/Magnetic Resonance Imaging in arrhythmic mitral valve prolapse. JAMA Cardiol. 2020;5(9):1000–5.

    Article  PubMed  Google Scholar 

  11. Sorensen J, Bergsten J, Baron T, Orndahl LH, Kero T, Bjerner T, et al. Myocardial external efficiency in asymptomatic severe primary mitral regurgitation using (11)C-Acetate PET. J Nucl Med. 2023;64(4):645–51.

    Article  CAS  PubMed  Google Scholar 

  12. Sorensen J, Stahle E, Langstrom B, Frostfeldt G, Wikstrom G, Hedenstierna G. Simple and accurate assessment of forward cardiac output by use of 1-(11)C-acetate PET verified in a pig model. J Nucl Med. 2003;44(7):1176–83.

    PubMed  Google Scholar 

  13. Harms HJ, Tolbod LP, Hansson NH, Kero T, Orndahl LH, Kim WY, et al. Automatic extraction of forward stroke volume using dynamic PET/CT: a dual-tracer and dual-scanner validation in patients with heart valve disease. EJNMMI Phys. 2015;2(1):25.

    Article  PubMed  PubMed Central  Google Scholar 

  14. Hansson NH, Tolbod L, Harms HJ, Wiggers H, Kim WY, Hansen E, et al. Evaluation of ECG-gated [(11)C]acetate PET for measuring left ventricular volumes, mass, and myocardial external efficiency. J Nucl Cardiol. 2016;23(4):670–9.

    Article  PubMed  Google Scholar 

  15. Danad I, Uusitalo V, Kero T, Saraste A, Raijmakers PG, Lammertsma AA, et al. Quantitative assessment of myocardial perfusion in the detection of significant coronary artery disease: cutoff values and diagnostic accuracy of quantitative [(15)O]H2O PET imaging. J Am Coll Cardiol. 2014;64(14):1464–75.

    Article  PubMed  Google Scholar 

  16. Danad I, Raijmakers PG, Driessen RS, Leipsic J, Raju R, Naoum C, et al. Comparison of coronary CT angiography, SPECT, PET, and Hybrid Imaging for diagnosis of ischemic heart Disease determined by fractional Flow Reserve. JAMA Cardiol. 2017;2(10):1100–7.

    Article  PubMed  PubMed Central  Google Scholar 

  17. Kajander S, Joutsiniemi E, Saraste M, Pietila M, Ukkonen H, Saraste A, et al. Cardiac positron emission tomography/computed tomography imaging accurately detects anatomically and functionally significant coronary artery disease. Circulation. 2010;122(6):603–13.

    Article  CAS  PubMed  Google Scholar 

  18. Nordstrom J, Kero T, Harms HJ, Widstrom C, Flachskampf FA, Sorensen J, et al. Calculation of left ventricular volumes and ejection fraction from dynamic cardiac-gated (15)O-water PET/CT: 5D-PET. EJNMMI Phys. 2017;4(1):26.

    Article  PubMed  PubMed Central  Google Scholar 

  19. Driessen RS, van Timmeren JE, Stuijfzand WJ, Rijnierse MT, Danad I, Raijmakers PG, et al. Measurement of LV volumes and function using Oxygen-15 water-gated PET and comparison with CMR Imaging. JACC Cardiovasc Imaging. 2016;9(12):1472–4.

    Article  PubMed  Google Scholar 

  20. Vanhove C, Franken PR, Defrise M, Momen A, Everaert H, Bossuyt A. Automatic determination of left ventricular ejection fraction from gated blood-pool tomography. J Nucl Med. 2001;42(3):401–7.

    CAS  PubMed  Google Scholar 

  21. Nordstrom J, Kvernby S, Kero T, Sorensen J, Harms HJ, Lubberink M. Left-ventricular volumes and ejection fraction from cardiac ECG-gated (15)O-water positron emission tomography compared to cardiac magnetic resonance imaging using simultaneous hybrid PET/MR. J Nucl Cardiol. 2023;30(4):1352–62.

    Article  PubMed  Google Scholar 

  22. Harms HJ, Knaapen P, de Haan S, Halbmeijer R, Lammertsma AA, Lubberink M. Automatic generation of absolute myocardial blood flow images using [15O]H2O and a clinical PET/CT scanner. Eur J Nucl Med Mol Imaging. 2011;38(5):930–9.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

The authors would like to thank the staff at Uppsala PET-centre for image data collection and production of radiotracers used in the study.

Funding

The study was supported by research grants 20130631 and 20190593 from The Swedish Heart-Lung Foundation.

Open access funding provided by Uppsala University.

Author information

Authors and Affiliations

Authors

Contributions

Study conception and design was developed by JoS and JeS. PET data collection and analysis were performed by JoS, HJH, JN, PS, EL, MJ, ML and JeS, and CMR analysis by TB, TK and JeS. JoS, ML and JeS performed statistical analysis. Inclusion of subjects was performed by JB, TB, FAF, LA and JeS. First draft of the manuscript was written by JoS. All authors read and commented on previous versions of the manuscript and approved the final manuscript.

Corresponding author

Correspondence to Jonathan Sigfridsson.

Ethics declarations

Ethics approval and consent to participate

The study was performed in accordance with the ethical standards as laid down in the 1964 Declaration of Helsinki and its later amendments. Imaging data were acquired in three studies, all approved by the Regional Ethical Review Board at Uppsala University (Dnr: 2012 − 543, 2020–07017, 2021–05230). All subjects provided written informed consent.

Consent for publication

Not applicable.

Competing interests

HJH, ML and JeS are working co-owners of Medtrace A/S. ML is an associate editor of the EJNMMI research.

Additional information

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Electronic supplementary material

Below is the link to the electronic supplementary material.

13550_2024_1150_MOESM1_ESM.docx

Supplementary Material 1: Fig. 1 – Example of gated PET images from the same patient. Left panel shows a15O-water gated scan with activity in the blood pool during the first 50 s of the scan. The right panel shows gating of corresponding11C-acetate retention images. Fig. 2 – Scatter plots and Bland-Altman plots comparing cardiac output (CO) calculated from left (LV) and right (RV) ventricular cavity input with 15O-water (WAT) (A, B) and 11C-acetate (ACE) (C, D). Dashed lines are lines of identity (A, C), dotted lines are limits of agreement (B, D) and solid lines represent linear regression and mean bias (B, D). Fig. 3 – Scatter plots of forward stroke volume (FSV) calculated with15O-water, 11C-acetate and cardiovascular magnetic resonance (CMR). Comparison between uncalibrated and calibrated PET values. Dashed lines are lines of identity.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Sigfridsson, J., Baron, T., Bergsten, J. et al. Quantitation of mitral regurgitation using positron emission tomography. EJNMMI Res 14, 85 (2024). https://doi.org/10.1186/s13550-024-01150-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13550-024-01150-1

Keywords