Skip to main content

Table 4 Comparison of cardiac function parameters obtained by IQ-SPECT and CMR using different reconstruction parameters

From: CMR validation of left ventricular volumes and ejection fraction measured by the IQ-SPECT system in patients with small heart size

 

Normal heart size

Small heart size

Normal heart size vs. CMR

Small heart size vs. CMR

3–10

 EDV

87.7 ± 14.9

46.2 ± 9.9

 < 0.001*

 < 0.001*

 ESV

43.8 ± 10.0

14.7 ± 6.5

 < 0.001*

 < 0.001*

 LVEF

50.6 ± 4.3

69.5 ± 8.6

0.117#

0.009*

5–12

 EDV

88.1 ± 14.8

49.4 ± 9.8

 < 0.001*

 < 0.001*

 ESV

44.5 ± 10.2

19.2 ± 6.9

 < 0.001*

 < 0.001*

 LVEF

50.0 ± 4.3

62.1 ± 7.8

0.051#

0.120#

5–15

 EDV

88.4 ± 14.9

51.1 ± 9.6

 < 0.001*

 < 0.001*

 ESV

45.6 ± 10.4

20.9 ± 7.0

 < 0.001*

 < 0.001*

 LVEF

49.0 ± 4.2

60.0 ± 7.1

0.013*

0.005*

8–12

 EDV

89.5 ± 15.2

52.3 ± 9.8

 < 0.001*

 < 0.001*

 ESV

47.4 ± 10.4

22.6 ± 7.2

0.001*

0.004*

 LVEF

47.4 ± 3.9

57.7 ± 7.3

 < 0.001*

 < 0.001*

8–15

 EDV

89.7 ± 14.8

53.1 ± 9.3

 < 0.001*

 < 0.001*

 ESV

49.3 ± 10.6

24.1 ± 7.1

 < 0.001*

0.025*

 LVEF

45.7 ± 5.0

55.3 ± 6.9

 < 0.001*

 < 0.001*

8–18

 EDV

90.9 ± 14.4

53.8 ± 9.9

 < 0.001*

 < 0.001*

 ESV

50.2 ± 10.7

24.8 ± 7.3

0.001*

0.054#

 LVEF

45.1 ± 4.3

54.6 ± 6.9

 < 0.001*

 < 0.001*

CMR

 EDV

129.7 ± 21.5

78.2 ± 18.8

  

 ESV

60.7 ± 11.7

28.0 ± 10.6

  

 LVEF

53.2 ± 5.8

64.6 ± 8.8

  
  1. *Represents P < 0.05; # represents P > 0.05, indicating a statistically nonsignificant difference
  2. EDV end-diastolic volume, ESV end-systolic volume, LVEF left ventricular ejection fraction, CMR cardiovascular magnetic resonance