Skip to main content

Table 2 The uncertainty u(D), Lin’s concordance correlation coefficient ρC and the normalized error EN for the simplified dosimetry protocols for the lesions

From: Optimization of the radiation dosimetry protocol in Lutetium-177-PSMA therapy: toward clinical implementation

Method

u(D) (%)

ρC

EN

1 h

24.3 ± 1.1

0.79

3.65 ± 1.86

24 h

21.2 ± 1.2

0.96

1.82 ± 1.32

48 h

17.5 ± 2.3

0.94

2.30 ± 1.96

72 h

15.8 ± 1.6

0.94

1.60 ± 1.79

168 h

20.1 ± 1.3

0.98

1.21 ± 0.94

1 h + 24 h

20.3 ± 16.1

0.66

7.24 ± 4.75

1 h + 48 h

18.8 ± 11.6

0.77

5.01 ± 3.69

1 h + 72 h

16.2 ± 6.2

0.84

3.60 ± 2.02

1 h + 168 h

13.8 ± 2.0

0.99

0.68 ± 0.61

24 h + 48 h

31.2 ± 24.9

0.77

3.69 ± 3.12

24 h + 72 h

18.7 ± 7.7

0.91

2.27 ± 1.97

24 h + 168 h

12.9 ± 2.3

0.99

0.65 ± 0.50

48 h + 72 h

24.7 ± 18.1

0.92

2.79 ± 1.89

48 h + 168 h

13.1 ± 3.0

0.99

0.65 ± 0.64

72 h + 168 h

13.4 ± 2.4

0.99

0.86 ± 0.48

  1. Mean ± SD values are given. Parameters that meet requirements are marked in bold. Protocols where all parameters meet requirements are marked in bolditalics