Table 4 Doses and activity concentrations at scan start times with acquisition times of the radiotracers on different PET/CT systems

From: Evaluation of image quality with four positron emitters and three preclinical PET/CT systems

PET/CT system Dose (MBq)/activity concentration (kBq/mL) at scan start time for each radionuclide
[11C]acetate [68Ga]chloride [18F]FDG [89Zr]oxalate
RAYCAN 3.69/167.73 3.72/169.55 3.70/168.18 3.68/167.27
Inveon 3.40*/154.55 3.73/169.55 3.76/170.91 3.68/167.27
Molecubes 3.69/167.73 3.72/169.09 3.70/168.18 3.68/167.27
  1. *Activity outside the specified limits of NEMA (3.70 MBq ± 5%)