Skip to main content

Table 2 Physiological, signal-to-noise ratio, and all lesion SUV parameters

From: Comparison between silicon photomultiplier-based and conventional PET/CT in patients with suspected lung cancer—a pilot study

SUV parameters (mean ± SD)

Discovery 690

Discovery MI

P value

Change (%)

Signal-to-noise ratio

10.9 ± 2.6

10.2 ± 2.2

0.53

− 6

Liver SUVmean

2.4 ± 0.5

2.2 ± 0.3

*0.01

− 10

Blood pool SUVmean

1.8 ± 0.3

1.7 ± 0.3

0.22

− 7

Primary lesions SUVmax

10.9 ± 8.2

10.3 ± 7.1

0.29

− 6

Primary lesions to blood pool SUVratio

6.6 ± 5.2

6.3 ± 4.6

0.84

− 5

Primary lesions to liver SUVratio

5.0 ± 3.9

4.9 ± 3.5

0.84

− 2

Lymph node SUVmax

6.4 ± 5.5

7.0 ± 5.2

*0.026

10

Lymph node to blood pool SUVratio

3.8 ± 3.5

4.2 ± 3.0

0.11

12

Lymph node to liver SUVratio

2.8 ± 2.6

3.2 ± 2.3

*0.021

17

  1. *Indicating a statistically significant difference