Table 2 Deviation between the activity concentration calculated from SPECT/CT images and the real activity concentration inside the bottle to simulate a kidney

From: Implementation of patient dosimetry in the clinical practice after targeted radiotherapy using [177Lu-[DOTA0, Tyr3]-octreotate

OSEM reconstruction corrections

Number of iterations (i) and subsets (ss)

Filter

Calculated activity concentration (MBq/ml)

Real activity concentration (MBq/ml)

Deviation (%)

AC + RR + SC

4i*5ss

None

0.506

0.537

− 5.88

4i*10ss

0.495

− 7.88

6i*10ss

0.502

− 6.55

8i*10ss

0.503

− 6.33

16i*10ss

0.502

− 6.55

32i*10ss

0.507

− 5.66

4i*5ss

Gaussian 0.11 cm

0.507

− 5.66

4i*10ss

0.494

− 88.0

6i*10ss

0.509

− 5.22

8i*10ss

0.507

− 5.66

16i*10ss

0.505

− 66.0

  1. AC attenuation correction, RR recovery resolution, SC scatter correction