Skip to main content

Table 1 ROI analysis of the emission image. Drawn ROIs nearby the headphones location and results

From: Alternative headphones for patient noise protection and communication in PET-MR studies of the brain

Phantom only (reference)

Phantom with headphones

Phantom with earphones

Mean (kBq/cc)

CV* (%)

Mean (kBq/cc)

CV* (%)

Deviation to reference

Mean (kBq/cc)

CV* (%)

Deviation to reference (%)

Ccentral ROI

24.6

37

24.5

37

0.4

24.2

36

1

Right ROI

27.2

33

24.7

35

10

27.0

33

0.7

Left ROI

23.2

44

21.0

45

10

23.0

44

0.9

  1. *Coefficient of variation