Table 2 Comparison of system characteristics across manufacturer PET/CT systems

From: Performance evaluation of the next generation solid-state digital photon counting PET/CT system

Manufacture

GE

GE

Philips

Philips

Philips

Siemens

Siemens

Toshiba

PET/CT model

Discovery MI (4-ring) [51]

Discovery

690 [43]

Vereos (this work)

Ingenuity TF [44]

Gemini T [42]

Biograph mCT flow [45]

Biograph mCT [46, 47]

Celesteion [48,49,50]

Photo detector

SiPM

PMT

SiPM

PMT

PMT

PMT

PMT

PMT

Number of detectors

9792

256

23,040

420

560

768

768

480

Scintillator

LYSO

LYSO

LYSO

LYSO

LYSO

LSO

LSO

LYSO

Number of crystals

19,584

13,824

23,040

28,336

28,336

32,448

32,448

30,720

Crystal size (mm3)

3.95 × 5.3 × 25

4.2 × 6.3 × 25

3.86 × 3.86 × 19

4 × 4 × 22

4 × 4 × 22

4 × 4 × 20

4 × 4 × 20

4 × 4 × 12

Ring diameter (cm)

74.4

81.0

76.4

90.0

90.3

84.2

84.2

88.0

Axial FOV (cm)

20.0

15.7

16.4

18.0

18.0

22.1

22.1

19.6

Plane spacing (mm)

n/a

n/a

1, 2, or 4

2 or 4

2 or 4

2

2

2

TOF Timing resolution (ps)

375

544

322

502

585

555

527

410

Sensitivity (cps/kBq)

13.7

7.4

5.7

7.3

6.6

9.6

9.7

4.0

Transverse resolution @ 1 cm (mm)

4.1

4.7

4.0

4.8

4.8

4.3

4.4

5.1

Transverse resolution @ 10 cm (mm)

5.0

5.1

4.4

5.1

5.2

4.9

4.9

5.1

Axial resolution @ 1 cm (mm)

4.5

4.7

4.0

4.7

4.8

4.3

4.4

5.0

Axial resolution @ 10 cm (mm)

6.0

5.6

4.8

5.2

4.8

5.9

5.7

5.4

Peak NECR (kcps @ kBq/mL)

193.4 @ 21.9

139.1 @ 29.0

171 @ 50.5

124.1 @ 20.3

125 @ 17.4

185 @ 29

156 @ 31.1

≥51 @ n/a

Energy resolution (%)

9.4

12.4

11.2

11.1

11.5

n/a

11.5

11.3

Scatter fraction at peak NECR (%)

40.6

37

30.8

36.7

27

33.4

32.7

42.7