| Aquiduo (Aquiduo) TOSHIBA | Biograph mCT 3 ring (Bio3R) SIEMENS | Biograph mCT Flow 4 ring (Bio4R) SIEMENS | Biograph 64 True Point (BioTP) SIEMENS | Celesteion PCA-9000A (Celesteion) TOSHIBA | Discovery IQ (DIQ) GE | Discovery ST Elite (DSTE) GE | Discovery STEP (DSTEP) GE | Discovery 600 Motion (D600) GE | Discovery 690 (D690) GE | GEMINI TF16 (GTF) Philips | Gemini GXL (GXL) Philips |
---|---|---|---|---|---|---|---|---|---|---|---|---|
Cross-calibration (background SUV) | 1.03 | 1.03 | 0.96 | 1.01 | 1.00 | 1.04 | 0.96 | 1.03 | 1.04 | 1.01 | 1.04 | 1.03 |
PET reconstruction | ||||||||||||
 Reconstruction | FORE OSEM | 3D-OSEM PSF TOF | 3D-OSEM PSF TOF | 3D-OSEM PSF | 3D-OSEM | 3D-OSEM PSF | 3D-OSEM | 3D-OSEM | 3D-OSEM | 3D-OSEM TOF | Full list mode TOF 3D-OSEM | LOR RAMLA |
 Iterations | 4 | 2 | 2 | 2 | 3 | 4 | 2 | 2 | 3 | 3 | 3 | 2 |
 Subsets | 14 | 21 | 21 | 21 | 10 | 12 | 28 | 21 | 16 | 8 | 33 | – |
 Smoothing | Gaussian | Gaussian | Gaussian | Gaussian | Gaussian | Gaussian | Gaussian | Gaussian | Gaussian | Gaussian | Smooth/sharp; smooth A | Smooth/sharp; normal |
 FWHM of filter (mm) | 8 | 6 | 4 | 4 | 6 | 5 | 6 | 5.14 | 4 | 4 |