Skip to main content

Table 3 Recovered activity in the liver background. Recovered activity concentrations are compared to measured values

From: Phantom validation of quantitative Y-90 PET/CT-based dosimetry in liver radioembolization

Liver Background

True activity concentration (MBq mL−1)

Recovered concentration (MBq mL−1) Average, [1st–3rd quartile] [% deviation]

Homogeneity Index

Reconstruction parameters

Scan Time (min)

1

0.89 ± 0.03

0.76 [0.430–0.920] [− 14.6%]

0.84

2i, 15s

15

4

0.41 ± 0.01

0.37 [0.160–0.460][− 9.8%]

0.94

2i, 15s

15

5

0.31 ± 0.01

0.26 [0.140–0.330] [− 16.1%]

0.91

2i, 15s

30

6

0.24 ± 0.01

0.19 [0.090–0.240] [− 20.8%]

0.90

2i, 15s

30

12

0.051 ± 0.002

0.021 [0.004–0.030][− 58.8%]

2.80

2i, 15s

30