Fig. 1 | EJNMMI Research

Fig. 1

From: Calibration of PET/CT scanners for multicenter studies on differentiated thyroid cancer with 124I

Fig. 1

Design of the phantom. All sizes are in millimeters. The height is lower than the smallest axial field of view of the PET/CT scanners included for calibration. The phantom consists of solid polyethylene (light gray) and contains six openings (indicated 1 to 6; 3 to 6 colored white). In each opening, a polyethylene socket can be placed tightly, enclosing an amber glass vial (dark gray, wall thickness about 1.5 mm) of 2 mL inner volume, filled with 124I activity (shaded). The position of the vials in the phantom is such that the 124I positrons (maximum range r) do not interfere between the vials and do not reach the outside of the phantom. The vials have a neck and screw cap, so the shape and height are an indication