Skip to main content

Table 1 Decrease in regional tracer uptake between scans without and with headphones

From: Impact of MR-safe headphones on PET attenuation in combined PET/MRI scans

Region

Decrease in uptake (%)

Cerebellum

10.8 ± 1.8

Mesial temporal lobe

8.8 ± 3.2

Temporal lobe

8.8 ± 4.3

Occipital lobe

7.5 ± 2.6

Fissura calcarina

6.8 ± 2.3

Basal ganglia

5.2 ± 2.7

Frontal lobe

4.2 ± 3.0

Gyrus cinguli

3.5 ± 2.7

Parietal lobe

3.3 ± 2.9

Central region

2.5 ± 2.2

  1. Values are given as mean ± standard deviation