Skip to main content

Table 2 Results of validation of HPLC versus ITLC (presented as 100% minus impurity B)

From: GMP-compliant 68Ga radiolabelling in a conventional small-scale radiopharmacy: a feasible approach for routine clinical use

Sample

Percentage of Ga-68-DOTA-NOC (HPLC)

Percentage of Ga-68-DOTA-NOC (ITLC 1)

Percentage of Ga-68-DOTA-NOC (ITLC 2)

Validation 1

98.2

98.3

98.5

Validation 2

98.1

98.0

98.3

Validation 3

98.7

99.2

98.5

Spike 5%

93.7

94.5

91.4

Spike 10%

89.5

89.9

89.6

Spike 15%

84.8

85.2

87.6