From: Factors affecting the stability and repeatability of gamma camera calibration for quantitative imaging applications based on a retrospective review of clinical data
Estimate
95% Confidence interval
Fixed effects
μ C F d A
0.1522
−0.0690 to 0.3734
β S E F
−0.0157
−0.1444 to 0.1759
βConstancy
−0.1317
−0.2935 to 0.0300
Random effects (weight of the parameter on the total variance)
σ C F d A 2
0.4443 (77.4%)
σ γ B k R 2
0.0510 (8.9%)
σ ε 2
0.0789 (13.7%)