Skip to main content

Table 2 Simulation measurements

From: Gating, enhanced gating, and beyond: information utilization strategies for motion management, applied to preclinical PET

Relative counts in image space

1

4

13

55

2,000

Maximum

 

Ungated

94

82

78

75

74

Gated

219

146

122

108

100

Gating+

95

125

115

108

100

Volume (70% maximum)

 

Ungated

57

144

177

194

199

Gated

6

17

49

84

99

Gating+

55

41

76

89

103

SUV (mean VOI/background)

 

Ungated

91

77

74

73

72

Gated

249

153

118

104

100

Gating+

92

122

110

105

100

FWHM

 

Ungated

144

174

183

187

188

Gated

38

68

90

98

100

Gating+

143

88

96

94

100

  1. Summary of lesion measurements from simulations of hot lesion against colder background with a range of random noise levels added (see Figure 5). Numbers represent average measurements from a single gate of 500 simulations for each noise scenario. All measurements are presented as a percentage of the corresponding true noise-free values. The standardized uptake value (SUV) was calculated from for 70% threshold volume.