Scan number | Events in scan (trues), ×10 8 | Median respiratory period (s) | Maximum COM displacement (mm) | Motion observed by all four independent observers | ||||
---|---|---|---|---|---|---|---|---|
Software | Hardware | Difference | Software | Hardware | Difference | |||
1 | 1.4 | 1.00 | 0.98 | 0.02 | 1.16 | 1.20 | −0.04 | Yes |
2 | 1.4 | 1.20 | 1.23 | −0.03 | 0.47 | 0.33 | 0.14 | Yes |
3 | 1.6 | 1.10 | 1.06 | 0.04 | 0.96 | 0.91 | 0.05 | Yes |
4 | 1.6 | 1.20 | 1.21 | −0.01 | 1.37 | 1.89 | −0.52 | Yes |
5 | 1.6 | 1.30 | 1.26 | 0.04 | 0.28 | 0.37 | −0.09 | Yes |
6 | 1.9 | 1.10 | 1.07 | 0.03 | 1.48 | 1.40 | 0.08 | Yes |
7 | 2.1 | 1.20 | 1.22 | −0.02 | 0.19 | 0.21 | −0.02 | Yes |
8 | 2.1 | 1.20 | 1.15 | 0.05 | 1.43 | 1.51 | −0.08 | Yes |
9 | 2.3 | 1.10 | 1.12 | −0.02 | 0.94 | 0.92 | 0.02 | Yes |
10 | 2.5 | 0.90 | 1.18 | −0.28 | 0.26 | 0.16 | 0.10 | No |
11 | 2.6 | 1.20 | 1.23 | −0.03 | 1.38 | 1.55 | −0.17 | Yes |
12 | 2.6 | 1.40 | 1.35 | 0.04 | 0.21 | 0.20 | 0.01 | Yes |
13 | 2.7 | 1.20 | 1.24 | −0.04 | 0.58 | 0.58 | −0.01 | Yes |
14 | 2.9 | 1.20 | 1.15 | 0.05 | 1.08 | 1.03 | 0.04 | Yes |
15 | 2.9 | 1.10 | 1.06 | 0.04 | 0.34 | 0.32 | 0.02 | Yes |
16 | 3.0 | 1.20 | 1.22 | −0.02 | 0.26 | 0.33 | −0.07 | Yes |
17 | 3.1 | 1.20 | 1.30 | −0.10 | 0.21 | 0.26 | −0.05 | No |
18 | 3.1 | 1.20 | 1.17 | 0.03 | 0.35 | 0.37 | −0.03 | Yes |
19 | 3.5 | 1.10 | 1.14 | −0.04 | 0.19 | 0.23 | −0.04 | Yes |
20 | 3.5 | 1.20 | 1.16 | 0.04 | 0.16 | 0.19 | −0.03 | Yes |
21 | 3.9 | 1.10 | 1.06 | 0.04 | 1.36 | 1.27 | 0.09 | Yes |
22 | 4.2 | 1.20 | 1.16 | 0.04 | 1.49 | 1.51 | −0.02 | Yes |